STEP 4. Insert a code chip
For measuring cholesterol (Lipid)
Coding
The code chip is included in the test strip package. The code chip provides the analyzer with
important information on the production-specific properties of the SD LipidoCare Professional
Lipid Profile Test Strip in order to measure your results accurately. Before you use the SD
LipidoCare Professional Analyzer with a new test strip for the first time, you should check that
the code numbers on the analyzer and the test kit are the same.
Code Setting
1) Make sure that the analyzer is turned off. Remove the old (or inserted) code chip from the
analyzer.
2) Insert a new code chip until it snaps into place. Then turn the analyzer on.
3) The 3-digit code number of the inserted code chip will appear with a flashing lipid test strip
symbol. This number must match the information provided in the lipid test strip package or
pouch. If it doesn’t match, repeat steps 1 and 2 above.
For measuring glucose
The blood glucose testing requires no coding. Therefore, a code chip is not needed for glucose
testing.
CAUTION:
The SD LipidoCare Professional
Lipid test strip
package includes a code chip with test strips.