User's Manual

Table Of Contents
Stage 10) Setting Hypo warning for glucose
In the 10th stage, you can set the hypo warning function for the glucose testing only.
1) After the printing setting, the display for setting the hypo warning will appear.
2) You can turn the hypo warning off or select the hypo level as 60, 70, or 80 mg/dL by
pressing the left or right button. Then save it by pressing the ON/OFF button.
NOTE:
Please consult your healthcare provider for an appropriate hypo level, from among 60,
70, or 80mg/dL, to be set for you.
If your glucose results are lower than a selected hypo result, the hypo symbol will
appear on the screen with a beep
sound. It is very important to manage your
hypoglycemia. Contact your healthcare provider immediately.
- 17 -