User's Manual
SD GlucoNavii® NFC BLOOD GLUCOSE MONITORING SYSTEM
Control Solution Test
29
t Shake the container, discard the rst drop of control
solution, and wipe o the tip to ensure a proper sample
and an accurate result.
t Store SD GlucoNavii® control solution tightly closed
at temperatures between 8-30°C (46-86°F). Do Not
refrigerate.
1. Performing Control Solution Test
You need the meter, a test strip, and control solution Level M or
Level H. The control level is printed on the test strip label.
For more information how to obtain SD GlucoNavii® control
solution, call at +82-31-300-0400.
t Do not swallow SD GlucoNavii® control
solution; it is not for human consumption.
t Do not apply SD GlucoNavii® control solution
to the skin or eyes as it may cause irritation.
CAUTION
A set of Level M and H control solutions is available
for purchase. To order control solutions, talk
to your pharmacist or medical surgical supply
dealers. Your meter is designed to recognize the
dierence between SD Glucose Control Solution
and blood. The meter automatically stores the test
results using a control solution, letting your review
them. But the meter does not include them in
averages.
NOTE