Service manual
REFRIGERATION SERVICE: R-404A
THIS ICE MACHINE USES R-404A (HP62)
REFRIGERANT AND POLYOLESTER
COMPRESSOR OIL.
DO NOT USE MINERAL OIL IN THIS
REFRIGERATION SYSTEM.
••HP62 is a "Near Azeotrope", and therefore
liquid charging is preferred.
••When the system is serviced, a special liquid
line dryer is required.
••Polyolester oil absorbs water very easily, and
therefore when the system is opened for
service, it must be re-sealed as soon as
possible (15 minutes maximum).
••Special leak detection equipment is required to
locate small refrigerant leaks. Usually a leak
detector capable of detecting a Halogenated
refrigerant or HFC-134A will work. Check with
the leak detector manufacturer if in doubt.
••As with any other refrigerant, do NOT mix HP62
with pressurized air when leak testing.
Pressure-Temperature Chart for HP62
    VAPOR       VAPOR 
TEMP.  PRESSURE  TEMP.  PRESSURE
(DEG F)  (PSIG)    (DEG F)    (PSIG)
-20 . . . . . .  17 70  . . . . . . 146
-18 . . . . . .  18 72  . . . . . . 150
-16 . . . . . .  20 74  . . . . . . 155
-14 . . . . . .  21 76  . . . . . . 161
-12 . . . . . .  23 78  . . . . . . 166
-10 . . . . . .  24 80  . . . . . . 171
 -8 . . . . . .  26 82  . . . . . . 177
 -6 . . . . . .  28 84  . . . . . . 182
 -4 . . . . . .  29 86  . . . . . . 188
 -2 . . . . . .  31 88  . . . . . . 194
 0  . .  . . . .  33 90  . . . . . . 200
 2  . .  . . . .  35 92  . . . . . . 206
 4  . .  . . . .  37 94  . . . . . . 212
 6  . .  . . . .  39 96  . . . . . . 219
 8  . .  . . . .  41 98  . . . . . . 225
10  . .  . . . .  43 100  . .  . . . 232
12  . .  . . . .  46 102  . .  . . . 239
14  . .  . . . .  48 104  . .  . . . 246
16  . .  . . . .  50 106  . .  . . . 253
18  . .  . . . .  53 108  . .  . . . 260
20  . .  . . . .  55 110  . .  . . . 268
22  . .  . . . .  58 112  . .  . . . 275
24  . .  . . . .  60 114  . .  . . . 283
26  . .  . . . .  63 116  . .  . . . 291
28  . .  . . . .  66 118  . .  . . . 299
30  . .  . . . .  69 120  . .  . . . 307
32  . .  . . . .  72 122  . .  . . . 316
34  . .  . . . .  75 124  . .  . . . 324
36  . .  . . . .  78 126  . .  . . . 333
38  . .  . . . .  81 128  . .  . . . 342
40  . .  . . . .  85 130  . .  . . . 351
42  . .  . . . .  88 132  . .  . . . 360
44  . .  . . . .  91 134  . .  . . . 370
46  . .  . . . .  95 136  . .  . . . 379
48  . .  . . . .  99 138  . .  . . . 389
50  . .  . . . .  102 140  . . .  . . 399
52  . .  . . . .  106 142  . . .  . . 409
54  . .  . . . .  110 144  . . .  . . 420
56  . .  . . . .  114 146  . . .  . . 430
58  . .  . . . .  118 148  . . .  . . 441
60  . .  . . . .  123 150  . . .  . . 452
62  . .  . . . .  127 152  . . .  . . 464
64  . .  . . . .  132 154  . . .  . . 475
66  . .  . . . .  136 156  . . .  . . 487
68  . .  . . . .  141 158  . . .  . . 499
FME800R
August 1993
Page 35