User manual

Windows Server Program
ComTec GmbH 5-21
[MAP0] 1A x 1B ; The following section concerns parameters of a MAP spectra
; named 1A x 1B, the number 0 is an index starting at 0 for the
; first calculated spectra
param=10000 ; Paramer definition (hex): the lower word defines the x-axis
; parameter (here 0 == ADC1A), the higher word the y axis (here
; 1 == ADC1B)
...
[TIM1] RTC x 1A ; The following section concerns parameters of a Time spectrum
; named RTC x 1A
param=101 ; Paramer definition (hex): the lower word defines the x-axis
; parameter (here 101 == RTC), the higher word the y axis (here
; 0 == ADC1A) The 1 msec Real Time (RT) has parameter value
; 100 (hex), the RTC 101 (hex)
The following commands define Replay settings and are not included in a MPA3.CNF file:
repluse=1 ; Set REPLAY mode ON (or OFF for repluse=0)
replmodif=1 ; Set Modified Settings for REPLAY to ON
; (or OFF for replmodif=0)
repllimit=1 ; Set REPLAY limited time Range to ON
; (repllimit=0 means complete listfile)
repltimefrom=0 ; Set Start time for Replay in listfile to 0 seconds
repltimepreset=60 ; Set Replay time preset in listfile to 60 seconds
replname=file ; Set list filename for REPLAY
replspeed=x ; Set Replay speed to x
...
A data header file with extension .MP contains a subset of above parameters and some
additional information typical for the special measurement. An example is:
ctm=80
dtm=80
sen=3
coi=3
aui=0
auo=0
auxsys=0
dac=0
diguse=0
digval=0
rtprena=0
rtpreset=1000.000
autoinc=0
savedata=0