Temperature calibration Reset(Calibration —— rCAL):
Press button (B),to jump to the next setting item.
Press button (A),to exit the setting mode, and display the operation temperature.
Press button (C) or (D) to enter the temperature calibration mode setting interface.
0:reset 1:1-points cal. 2:2-points cal. 3:3-points cal.
When the setting value is “0”, press button (B) to reset the previous calibration data.
Temperature calibration point 1:
The interface for viewing the value before correction:
Press button (B),
to jump to the next CLP2 item.
Press button (A),to exit the setting mode, and display the operation temperature.
Press button (C) or (D) to enter the interface for viewing the value.
Press button (B),
to jump to the temperature calibration point 1 deviation value
interface .
Press button (A),to exit the setting mode, and display the operation temperature.
Deviation value interface
:
Press button (B),
to jump to the next CLP2 item.
Press button (A),to exit the setting mode, and display the operation temperature.
Press button (C) or (D) to enter the temperature calibration point 1 deviation value input
interface
.
Press button (B),save the deviation value and
jump to the next CLP2 item.
Press button (A),to exit the setting mode, and display the operation temperature.
The device can be calibrated a total of 3 points.