User Manual

13
Rotary Evaporator
Release Lift-up/Lift-down key, the lift stops.
Push the speed key and then rotate control knob to set
the rated speed.
Push the temperature key and then rotate control knob
to set the rated temperature.
Push the timer key and then rotate control knob to set
the rated time.
Deactivate the timer by setting the target value to 0.
Press the Int key and then rotate control knob to set the
rated interval value.
Deactivate the interval by setting the target value to
0.
Press the control knob, switch ON.
If these operations above are normal, the device is ready to
operate. If not, the device may be damaged during
transportation, please contact manufacturer/supplier for
technical support.
Warning !
Forbid to transfer the flask when the instrument
working.
6. Faults
Instruments can’t be power ON
- Check whether the power line is unplugged
- Check whether the fuse is broken or loose
Fault in power on self test
- Switch OFF the unit, then switch ON and reset
the instruments to factory default setting.
If these faults are not resolved, please contact
manufacturer/supplier.