Owner's manual

11
Digital 7” Square(Hotplate)Magnetic Stirrer
The unit can be controlled from an external PC (using the
dedicated software) via the RS232 C serial interface fitted to
the unit. Data communication from laboratory instrument to
computer is only possible on demand of the computer.
•
The functions of the interface lines between laboratory
instrument and automation system are selected from
the specified signals of the EIA-standard RS232C,
corresponding with DIN66020 Part 1. The allotment of
the bushing can be taken from Figure 7.
•
Transmission method: Asynchronous signal transmission
in start-stop-operation.
•
Mode of transmission:Fully Duplex.1start bit; 7 character
bits; 1stop bit.
•
Transmission speed: 9600 bit/s.
•
Start remote control knob LED display Remote.
Note:
F o r b i d t o in s e r t or r e m o v e th e R S 2 3 2 C
communication line when switch on!
10.Faults
•
Instruments can not be power ON
-
Check whether the power cable is plugged in
-
Check whether the fuse is broken or loose
•
Fault in power on self test
-
Switch OFF the unit, then switch ON and reset the
instruments to factory default setting.
•
Temperature can switched on not reach set point
-
Check whether the safety temperature value is set too
low.
•
Stir speed can not reach set point
-
Check whether the stirring bar breakaway monitoring
is ON.
-
Excessive medium viscosity may cause abnormal
speed reduction of the motor.
•
Heating or stirring function do not start when knob is
pressed.
-
Check whether the unit is in the mode setting program,
-
Exit from the mode setting and restart the heating/
stirring functions.
•
Unit do not power down when switched off.
-
Check if the residual heat warning function is still ON
and hotplate temperature is above 50(the LCD still
works andHotflashes).
If these faults are not resolved,please set the instruments
to factory default setting or take the unit to technical service
center.