In the B mode, if the external temperature sensor is not
plugged in, Probe will flash. If plugged in, Probe
will be shown on the digital display to indicate that the
sensor is operating. The external temperature sensors
value and actual temperature are displayed.
Compared with the temperature control of the hotplate, the
external temperature sensor can control the mediums
temperature more precisely.
Warning!
Forbid to Plug in/pull out external temperature
sensor in heating.
8.2 Residual heat warningHOT
In the A or B mode, users can set the residual heat warning
function, as follows:
Press continuously the function setting key till the LCD
displays rES.
Press the temperature regulation key to switch on or off
the residual heat warning function, and the LCD displays
On or OFF.
Press continuously the function setting key till Set
disappear.
The residual heat warning function has been switched on
or off.
In order to prevent the risk of burns from a hotplate, MS-H-
Pro has a residual heat warning function. When the heating
function is switched off and the hotplate temperature is
still above 50, Hot will flash to warn that there is a
hazard of burns from the hotplate. Wh
en the heating plate
temperature drops to below 50, the unit will automatically
switch off. If users want to turn off the LCD immediately,
just pull out the plug directly. When the plug is pulled out,
the residual heat warning function can not be run.
8.3 Setting the safety temperature
In the A or B mode, the safety temperature can be set, as
follows:
Press continuously the function setting key till the LCD
displays SAFE.
Adjust the temperature setting key to set the safety
temperature from 100 to 350.
Press continuously the function setting key till Set
disappear.
The value of the safety temperature has been stored and
setting is completed.