Instruction Manual

18
11.2 Tubes
11.2.1 Cleaning and sterilizing tubes
OApplicable XInapplicable
Conditions Materials PA PC PP
Cleaning
Cleaning fluids
AcidicpH5 or lower X X X
Acidichigher than pH5 O O O
Alkalinehigher than pH9 O X O
AlkalinepH9 or lower O O O
NeutralpH7 O O O
Warm water(up to 70 ) O O O
Ultrasonic cleaning Neutral detergentpH7 O O O
Sterilization
Autoclaving
1150.7kg/cm
2
30minutes O O O
1211.0kg/cm
2
20 minutes X O O
1261.4kg/cm
2
15 minutes X X X
Boiling 15 to 30 minutes O O O
Ultraviolet sterilization 200-300nm X X X
Gas sterilization Ethylene oxide O X O
Formaldehyde O O O
PA: Polyallomer PC: Polycarbonate PP: Polypropylene
Table 11.2 Cleaning and sterilizing conditions for tubes
11.2.2 Cleaning PC tubes
PC material is low in chemical resistance against alkaline solutions. Avoid using neutral detergents with
pH higher than 9. Note that pH of some neutral detergents are still higher than 9 even if diluted according
to the manufacturers instructions. Use detergent with its pH between 7 and 9.
11.2.3 Autoclaving PAPC and PP tubes
PA begins softening at about 120 , PC and PP at about 130℃℃. Autoclave PA tubes at 1150.7kg/cm
2
for 30 minutes, PC and PP tubes at 1210.1kg/cm
2
for 20 minutes. If a certain temperature is exceeded,
the tubes may be deformed.
When use a sterilizing chamber, please operate as follows:
(1) Place tubes in vertical position, mouths upward. If tubes are placed sideways, they may deform into an
oval shape due to gravity.
(2) Remove locking nut and lid to prevent from deformation or rupture.
(3) Wait until the sterilizing chamber cools down to the room temperature before removing tubes.