Manual
16
11.2.4 Conditions and life expectancy of tubes
The life expectancy of plastic tubes depends on the characteristics of samples, speed of the rotor used,
temperature applied and so on. When the plastic tubes are used for ordinary aqueous samples (pH between
5.0 and 9.0), their life expectancies are defined as follows.
Be operated at the maximum speed:
High quality tubes (PA、PC、PP): 30-50 operations
Ordinary tubes(PA、PC、PP): around 10 operations(Using in low speed can extend the tube life).
Life expectancy of tubes also depends on the pretreatment conditions such as cleaning and sterilization,
lifetime can be cut down.
Notice: Do not use damaged or cracked tubes.
12. Calculate RCF
An RCF can be determined with the following calculation formula.
RCF=1.118×r×n
2
×10
-5
r—rotating radius, unit: cm; n—rotating speed, unit: rpm
13. Ordering information
Cat. No. Model Descriptions
913023419999 DM0412
Clinical Centrifuge, with A12-10P rotor, US plug, 110V-240V,
50Hz/60Hz
913123419999 DM0412
Clinical Centrifuge, with A12-10P rotor, Cn plug, 110V-240V,
50Hz/60Hz
913223419999 DM0412
Clinical Centrifuge, with A12-10P rotor, Euro plug, 100V-240V,
50Hz/60Hz
913323419999 DM0412
Clinical Centrifuge, with A12-10P rotor, UK plug, 110V-240V,
50Hz/60Hz
Accessories
19200316 A12-10P Plastic rotor, 4500rpm, 15ml*8, used with DM0412
19200317 A10P15 Rotor adapter, used with A12-10P, 12pcs/pk
19200318 A10P15 Rotor adapter mat, used with A10P15, 12pcs/pk