D-160 HOMOGENIZER ____________________________________________________________________________ CE/ cULus - Declaration of Conformity EC-Declaration of Conformity
Product Description The D-160 hand held Disperser is a high speed dispersing instrument. It is based on the Rotor/Stator Technology. The shaft and rotor/stator can be disassembled for easy cleaning. You will receive the dispersing tools completely assembled for immediate use with your disperser. Should you dismantle these (e.g. for cleaning) please refer to page 4.
When cleaning, remove the power cord from the power source. Never touch the spinning rotor, or shaft, or motor side coupling parts. Do not operate if damaged in any manner. Return unit to our service centre for examination and repair. Switch off the unit before changing the dispersing element. Only suitable dispersing shafts from Scilogex must be used. The ventilation slots of the drive must not be obstructed. Never let the aggregate touch the bottom of the vessel.
Turn the motor on with the stator/shaft tube removed it may bend the axle/rotor. Always, both axle/rotor and stator/shaft tube must be attached.
must press the black button to lock it Put the shaft tube onto the axle and screw clockwise onto the housing! Press the black locking button and screw it fully onto the drive! When assembling the dispersing tool, always make sure that the PTFE bearing is fixed correctly and cooled by the working medium. Otherwise, it may cause serious defects to the whole dispersing tool. Examine the PTFE bearings regularly. Treat the dispersing tools carefully as they are the hearts of your dispersing unit.
that dissolves the residue, but does not harm the Teflon bearing and the steel. For proper cleaning, the dispersing element must be disassembled. Proper care and cleaning of the equipment will ensure a longer and better use of the equipment. Sterilization Chemical processes: Germicidal solutions (formalin, phenol, alcohol etc) can disinfect in most cases. However, residues of the genome must subsequently be removed with sterilized water.
1= 8,000 rpm 2= 12,000 rpm 3= 15,000 rpm 4= 18,000 rpm 5= 22,000 rpm 6= 30,000 rpm ______________________________________________________________________ For an emergency stop, press the large button on the drive. ______________________________________________________________________ Maintenance of Motor The motor does not need any maintenance and there are no parts inside which can be repaired by the user. The only parts which are excluded from this are the carbon brushes.
Relative Humidity Permissible Period of Operation Protection Dimensions WxDxH (drive) 85% 100% IP 30 46mm x 55mm x 230mm Warranty This instrument has a warranty of 24 months from date of purchase which covers material and workmanship. Scilogex, LLC will repair or replace free of charge the defect parts which were found defective after an inspection finds that the defect is due to materials or workmanship.