Models S310 & S310D Laser Power and Energy Meters Setup and Operating Procedures Serial Number___________________ P/N 10190M
Thank you for choosing a Scientech laser power and energy meter. Scientech, an ISO 9001 registered company, and our employees are pleased to provide you with a product designed for years of reliable service. Please read this manual completely before using your indicator. This information will enable you to fully utilize the equipment and should be located nearby for reference. The indicator is intended to be used only in the manner outlined in this manual. Misuse of the equipment may cause product failure.
Table of Contents: Detector Operating Parameters: ................................................................................................................................................................. 2 CE Mark Certification: .............................................................................................................................................................................. 4 Environmental Requirements:.......................................................................
3. To Zero the Analog Needle (does not apply to the S310D): ...................................................................................................... 28 4. To Connect an Ultra Detector: ................................................................................................................................................... 28 5. To Select a Group:.............................................................................................................................................
VECTOR™ S310 INDICATOR SPECIFICATIONS: Model Display S310 S310D 4 Digit LCD with Selectable 4 Digit LCD Analog Meter Movement 10.00 m, 100.0 m, 1.000, 10.00, AUTO (Watts only) 300.0 m, 3.000, 30.00, AUTO (Watts only) 3.000 m, 30.00 m, 300.0 m, 3.000, 30.00, AUTO 3.000 µ*, 30.00 µ*, 300.0 µ, 3.000 m, 30.00 m, AUTO 150.
VECTOR™ PYROELECTRIC DETECTOR SPECIFICATIONS: Model Maximum Beam Diameter Spectral Response Maximum Average Power Minimum Energy Noise Equivalent Energy Maximum Energy Density Accuracy Output Sensitivity Maximum Repetition Rate Maximum Pulse Duration Dimensions D x L - inches cm Weight - pounds/kgs Indicator Compatibility P 25 PHF 25 25mm 25mm .193-26µm PHD 25 PHDX 25 PHDX25UV SP 25 SPHF 25 SPHD 25 25mm 7mm 7mm 25mm 25mm 25mm .193-10.6µm .4-2µm .193-2µm .193-10.6µm .
VECTOR HR PYROELECTRIC DETECTOR SPECIFICATIONS: Model Active Diameter Spectral Response Voltage Response S, I L Electrical Decay Time (RC Time Constant) S I L Noise Equivalent Energy S, I L Minimum Energy S, I L Rep Rate (max) S I L Maximum Pulse Width (For Calibrated Response) S I L Maximum Voltage Output Maximum Average Power Accuracy PHF 02 2 mm PHF 05 5 mm PHF 09 9 mm .193 - 26 µm 15 V/mJ 0.15 V/mJ 2.5 V/mJ 0.025 V/mJ 1 V/mJ 0.01 V/mJ 0.05 msec 0.5 msec 2.5 msec 0.05 msec 0.5 msec 2.5 msec 0.
ASTRAL™ CALORIMETER SPECIFICATIONS: Model Type of Absorber Maximum Beam Diameter Spectral Response Average Power Maximum Average Power Minimum Noise Level Maximum Power Density Maximum Peak Power Density Maximum Single Pulse Energy Maximum Energy Density Precision Accuracy Response Time Dimensions DxL - inches cm Weight - pounds/kg Indicator Compatibility Model Type Absorber Maximum Beam Diameter Spectral Response Average Power Maximum Average Power Minimum Noise Level Maximum Power Density Maximum Peak Pow
LARGE APERTURE (100MM & 200MM) CALORIMETER SPECIFICATIONS: Model Type of Absorber Aperture Diameter Minimum Beam Diameter Spectral Response Maximum Average Power Minimum Average Power Noise Level Maximum Power Density Maximum Peak Power Density Maximum Single Pulse Energy Maximum Energy Density Precision Accuracy Response Time Dimensions DxL - inches/cm Weight - pounds/kgs Indicator Compatibility * This is a segmented absorber Note 1: 380401 Note 2: 384UV5, 388UV5 Note 3: 380801 Note 4: 380401 Note 5: 384UV
ULTRA™ CALORIMETER SPECIFICATIONS: Model Type of Absorber Maximum Beam Diameter Spectral Response Maximum Average Power Minimum Average Power Noise Level UC150 Surface 25mm .25 - 35µm UC150HD Surface 25mm .193 - 12µm Maximum Power Density 200W/cm2 1.5kW/cm2 1.
QUICK SETUP: Note: For detailed instructions for each type of detector, refer to the Operating Procedures section. 1. Turn On the Meter: Note: For the most accurate measurements possible, the S310 should be turned on and warmed up for 30 minutes. Note: For information on Group Settings, refer to the Group Settings section. Press the ON switch located in the rear panel of the meter. The S310 will immediately turn on with its operational state based on the last used detector Group Setting.
5. Select a Range: On the first press of the RANGE button, the analog scale (if active) will disappear. Then briefly press the RANGE button each time another range is desired. Press the SELECT button when the range you desire appears in the display (failure to press the SELECT button within 3 seconds will activate the range shown in the display and return the S310 to its operation state). 6. Select a Mode: The MODE button allows selection of the type of measurement to be made (watts, joules, etc).
2. Group Settings for Astral or Large Aperture Calorimeters: Note: The factory default Group for Astral or Large Aperture Calorimeters is Group #1. Note: Pressing the CANCEL button at any time during this process will terminate this process with no changes made. Note: Holding down the SELECT button for 3 seconds at any time during this process will save any changes made up to that point and return the S310 to its operational state for the Group selected.
To set the time delay between pulses use the count up (RANGE) and count down buttons (MODE) buttons to enter the time in seconds. Press the SELECT button after the time has been entered to save the setting.
3. Group Settings for Vector Pyroelectric Detectors and Vector HR Pyroelectric Detectors: Note: The factory default Group for Vector Pyroelectric Detectors is Group #2. Note: The factory default Group for Vector HR Pyroelectric Detectors is Group #3. Note: Pressing the CANCEL button at any time during this process will terminate this process with no changes made.
H. The "At" (attenuation) annunciator now appears in the display. This allows you to enter a value from 0.0001 to 9999 which will become a multiplier of the actual displayed value. If no attenuator or correction factor is to be used in conjunction with the pyroelectric detector then the attenuation factor must be set to 1.000. The attenuation factors of optics can be entered into the H410 so the displayed value will automatically compensate for the amount of attenuation.
C. To select a Group, press the SELECT button when the Group you desire appears in the display (failure to press the SELECT button within 3 seconds will activate the Group shown in the display and return the S310 to its operational state). D. Press the SETUP button until the "ULtr" annunciator appears in the display. Immediately press the SELECT button to select the Ultra detector configuration program. E. The "SET CAL" (calibration) annunciator now appears in the display.
OPERATING PROCEDURES: USING THE S310 WITH VECTOR PYROELECTRIC DETECTORS: Pyroelectric detector models P25, P50, SP25, SP50, P05, and P09 are coated with a special black absorbing material which provides a very flat spectral response over a broad wavelength band. Pyroelectric detector models PHD25, PHDX25, PHDX25UV, PHD50, PHDX50, PHDX50UV, SPHD25, and SPHD50 are coated with a special high damage absorbing material which provides absorption over a broad wavelength band.
3. To Zero the Analog Needle (does not apply to the S310D): Refer to Figure 3. The black slotted button located just below the display allows screwdriver adjustment to set the analog needle to zero. This adjustment should be made before connecting a pyroelectric detector. 4. To Connect a Pyroelectric Detector: Note: Only one detector should be plugged in at any time. Refer to the drawing of the S310 rear panel in Figure 2.
7. To Select a Mode: Available pyroelectric modes are: Energy (J), Avg. Energy (J AVG), Average Power (W AVG), and Volts (V). A. To Measure Energy (J): Note: The update rate of the display is 20Hz. i. Press the MODE button (repeatedly, if necessary) until the “J” annunciator appears in the display. ii. Press the SELECT button (or wait 3 seconds) and the joules mode will automatically be activated. iii. The energy level of each laser pulse will be displayed on the LCD. B.
8. To Measure a Statistical Run of Energy Pulses: Note: Do not use AUTO range when making a statistical run. Note: Do not go from the Average Power Mode to Stats since the range will be too high. Select the range manually. Note: When using the PHF09 pyroelectric detector in the long pulse setting, a manual range must be selected when running statistics. Note: Each time a new stats run begins, data from the previous run is lost.
10. To Perform a Transfer Calibration: You can transfer a calibration from a calorimeter to a Vector pyroelectric detector using the Transfer Calibration function of the S310. To do this, please refer to the Group Settings for Vector Pyroelectric Detectors, paragraph #G. You must follow the Group Settings steps to get into the Transfer Calibration function.
CORRECTING PYROELECTRIC DETECTOR OPERATING PARAMETERS FOR USE AT DIFFERENT WAVELENGTHS: NOTE: Due to variability in the manufacturing process the absorption characteristics of the HD and P model detectors can vary in the UV region (190 to 400nm). Scientech recommends optical calibration at 266nm if the detector is to be used in the UV region instead of relying on an absorption rate in the chart at the end of this manual.
2. To Turn the Analog Needle On or Off (does not apply to the S310D): To turn the analog meter on or off, press and keep holding the RANGE button down. Then release after the meter appears or disappears. 3. To Zero the Analog Needle (does not apply to the S310D): Refer to Figure 3. The black slotted button located just below the display allows screwdriver adjustment to set the analog needle to zero. This adjustment should be made before connecting a calorimeter. 4.
7. To Select a Mode: The modes available for the calorimeters are: Average Power (W AVG) or Energy (J). A. To Measure Average Power (W AVG): Note: The speed-up circuit (differentiator) in watts mode is controlled by software and was adjusted at the factory to accommodate the type of calorimeter being used. You can change the speed-up setting to your preference.
The statistics mode will collect data on a pulse population of up to 1000 pulses. At your prompting, the indicator will display the number of pulses delivered, average energy, minimum energy, maximum energy, standard deviation, and coefficient of variation. When the statistics mode is selected, the energy mode is automatically activated regardless of the mode previously selected. Select the appropriate range for the pulse energy level to be measured.
The following are viable external triggers. The following pulses will not trigger the S310. CORRECTING ASTRAL SERIES HD CALORIMETERS OPERATING PARAMETERS FOR USE AT DIFFERENT WAVELENGTHS: NOTE: Due to variability in the manufacturing process the absorption characteristics of the HD and P model detectors can vary in the UV region (190 to 400nm).
USING THE S310 WITH ULTRA SERIES DETECTORS: The Ultra detector selected needs to be the appropriate model for the planned laser measurements. Please familiarize yourself with the operating specifications which are given in the front of this manual. Note: The default Group Setting for Ultra Detectors is Group #4. Either select Group #4 or configure another Group for Ultra Detectors and select that Group. 1.
CORRECTING ULTRA SERIES HD CALORIMETERS OPERATING PARAMETERS FOR USE AT DIFFERENT WAVELENGTHS: NOTE: Due to variability in the manufacturing process the absorption characteristics of the HD and P model detectors can vary in the UV region (190 to 400nm). Scientech recommends optical calibration at 266nm if the detector is to be used in the UV region instead of relying on an absorption rate in the chart at the end of this manual.
1. Calorimeter Watts Mode: Compare voltage to the watt reading displayed on the S310 to determine V/W constant to use when looking at the analog output. In power (watts) mode, the analog output is nominally 3 V full scale for each range. Therefore, to calculate watts from the analog output voltage: W = V/S where: V = analog output voltage (V) S = analog output sensitivity (V/W) ≈ 3 V/range (W) (for example, on the 10 mW range, S ≈ 3 V/.010 W = 300 V/W) 2.
3. Pyroelectric Watts and Joules Modes: Compare peak voltage to energy (joules) or power (watts) reading displayed on the S310 to determine V/J or V/W constant to use when looking at the analog output. In power (watts) and energy (joules) mode, the peak voltage of a laser pulse is directly proportional to the power energy level: W = Vpk/S where: Vpk = peak voltage from the analog output S = analog output sensitivity (V/W) ≈ 3 V/range (W) (for example, on the 3 mW range, S ≈ 3 V/.
CALIBRATION OF ASTRAL CALORIMETERS USING ELECTRIC SUBSTITUTION HEATING: For Astral calorimeters the electric substitution heating option must be ordered and installed at the factory when the calorimeter is purchased. It can not be retrofitted to a calorimeter at a later time. To calibrate using electric substitution heating proceed as follows. Figure 5 A. Remove the screws holding the calorimeter's ID tag and remove the plate to expose the circuit board as shown in Figure 5. B.
CALIBRATION OF LARGE APERTURE CALORIMETERS USING ELECTRIC SUBSTITUTION HEATING: Electrical substitution heating is a standard feature of large aperture calorimeters. Figure 6 Calibration with an Interface Module and S310 Indicator: A. Connect a DVM to the white jacks of the calorimeter. Refer to Figure 6. B. Measure the resistance of the substitution heater making sure to subtract the resistance of the patch cables from the total resistance measurement.
where: V = voltage applied to the heater coil Rc = substitution heater resistance from step B C = Cal coefficient 360401 = 1.018 360801 = 1.000 380401 = 0.974 380801 = 1.008 380402 = 1.024 380802 = 1.008 384UV5 = 1.021 388UV5 = 1.002 W = desired laser power in watts D. Connect the DVM to the calorimeter’s DIN connector. E. Apply the calculated voltage (V) to the electrical substitution heater. F. Record the voltage reading of the DVM (Vc). G.
Operation of Astral and Large Aperture Calorimeters with a Digital Volt Meter: Note: Whenever a large aperture calorimeter is used without an indicator the interface module is not used . The calorimeters may be used with any digital volt meter (DVM) capable of reading 5 volts full scale. Connect the output of the calorimeter to the DVM. A. Select the DC volts mode. B. Direct the laser beam on to the absorbing surface of the calorimeter. C.
The error, in theory, is only dependent upon the value of ∑dVi, that is the cumulative random error of Vi. This number should approach zero if data is carefully taken. The accuracy is also increased if the time interval, dt, is minimized. Numerical integration can yield accurate results, but is a tedious task.
Remote Interface Language: The remote interface language for the S310 is compatible with Std-IEEE488.2 and it also works with the RS232 remote interface. 1. RS-232 Connector: The RS-232 connector is a 9 pin subminiature connector on the instrument rear panel. The pin out and pin descriptions are shown below: Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Pin 7 Pin 8 Pin 9 Unused Data In (RXD) Data Out (TXD) DTR (is generated) Ground DSR (is ignored) RTS (is marking) CTS (is evaluated if requested) Unused 2.
4. Remote Interface Language Common Commands and Queries: Most commands and queries may be used with either the RS232 interface or the IEEE488 interface. Some commands, however, work only with the IEEE488 interface. Some commands require that a password has been entered. A few commands are archaic, but were left in because no harm was done, and they may be needed again if multi-channel units are constructed again.
RANGE_|auto Sets the range to the decimal number supplied, or to the closest legal range allowed for the current detector and configuration. If the word auto is supplied, this command activates the autorange feature if it is supported for the current detector and configuration. RANGE? The RANGE query returns the decimal number between 1 and the maximum range of the instrument. The meaning of the number varies, depending upon the type of detector, and the configuration of the instrument.
DUMP? If statistics have been collected from the front panel, or by the COL or COL? commands, this command returns all of the data values in the collection. If no statistics have been collected, it does nothing. BAUD_ If the takes any of the values 300, 1200, 2400, 9600, or 19200; or any of their abbreviations 3, 12, 24, 96, 192, this command sets the RS232 baud rate to the corresponding value.
*RCL_ This is the IEEE488.2 common command. It takes the single numeric argument [1-4], and restores the instrument configuration to that configuration stored in the corresponding save area. It also places the IEEE interface into the OCIS and OQIS states. COUNT_ Sets the number of pulses making up each average in average energy mode. COUNT? Returns the number of pulses making up each average in average energy mode.
5. IEEE488 Specific Commands: These commands may be used only from the IEEE488.2 interface. They are all members of the collection of the so-called "common Commands" described in the IEEE488.2 standard. *CLS Clears the Standard Event Status Register and forces the device into Operation Complete Command Idle state. *ESR? Returns a decimal number which is the value of the Standard Event Status Register. Reading the register clears it.
RETURNED GOODS PROCEDURE: Should it become necessary to return any item to Scientech for any reason, please contact our Product Service Department at (800)525-0522 or (303)444-1361 or Fax (303)444-9229 or email inst@scientech-inc.com. When you call, please be ready to provide model number, serial number, and a description of the problem. Frequently we can provide self-help information which will eliminate the need for returning the unit(s).
HD CALORIMETER AND PYROELECTRIC ABSORPTION VS. WAVELENGTH: Use this table for all HD calorimeters and HD pyroelectric detectors. Wavelength µm 0.200 0.210 0.220 0.230 0.240 0.250 0.255 0.260 0.265 0.270 0.275 0.280 0.285 0.290 0.295 0.300 0.305 0.310 0.315 0.320 0.325 0.330 0.335 0.340 0.345 0.350 0.355 0.360 0.365 0.370 0.375 0.380 0.385 0.390 0.395 0.400 0.405 0.410 0.415 0.420 0.425 0.430 0.435 0.440 0.445 0.450 0.455 0.460 0.465 0.470 0.475 Absorption % 93.66 93.78 92.18 90.44 88.79 90.07 90.42 90.
Wavelength µm 1.245 1.250 1.255 1.260 1.265 1.270 1.275 1.280 1.285 1.290 1.295 1.300 1.305 1.310 1.315 1.320 1.325 1.330 1.335 1.340 1.345 1.350 1.355 1.360 1.365 1.370 1.375 1.380 1.385 1.390 1.395 1.400 1.405 1.410 1.415 1.420 1.425 1.430 1.435 1.440 1.445 1.450 1.455 1.460 1.465 1.470 1.475 1.480 1.485 1.490 1.495 1.500 1.505 Absorption % 91.78 91.79 91.76 91.75 91.76 91.79 91.75 91.74 91.74 91.74 91.71 91.72 91.70 91.70 91.73 91.78 91.78 91.79 91.76 91.77 91.82 91.94 92.17 92.17 92.17 92.15 92.12 92.
Wavelength µm 2.129 2.130 2.132 2.135 2.136 2.139 2.140 2.143 2.145 2.146 2.150 2.153 2.155 2.157 2.160 2.161 2.164 2.165 2.168 2.170 2.171 2.175 2.179 2.180 2.182 2.185 2.186 2.190 2.193 2.195 2.197 2.200 2.201 2.205 2.208 2.210 2.212 2.215 2.216 2.220 2.224 2.225 2.227 2.230 2.231 2.235 2.239 2.240 2.243 2.245 2.247 2.250 2.251 Absorption % 90.76 90.41 90.05 90.41 90.43 90.84 90.48 90.46 91.65 90.04 90.45 90.44 90.42 90.39 90.36 90.69 90.76 90.43 90.44 90.44 90.41 90.38 90.38 90.37 90.40 90.44 90.42 90.
Wavelength µm 2.855 2.862 2.868 2.874 2.881 2.887 2.894 2.900 2.907 2.913 2.920 2.926 2.933 2.940 2.946 2.953 2.960 2.966 2.973 2.980 2.987 2.994 3.001 3.008 3.015 3.022 3.029 3.036 3.043 3.050 3.057 3.065 3.072 3.079 3.087 3.094 3.101 3.109 3.116 3.124 3.131 3.139 3.146 3.154 3.162 3.170 3.177 3.185 3.193 3.201 3.209 3.217 3.225 Absorption % 89.26 89.24 89.25 89.23 89.24 89.26 89.24 89.19 89.22 89.20 89.22 89.24 89.20 89.21 89.17 89.17 89.17 89.17 89.15 89.12 89.14 89.17 89.14 89.11 89.09 89.08 89.08 89.
Wavelength Absorption Wavelength Absorption Wavelength Absorption Wavelength µm % µm % µm % µm 5.357 86.95 6.682 86.05 8.879 84.32 13.09 5.379 86.95 6.717 86.06 8.940 84.27 13.23 5.401 86.94 6.752 86.07 9.002 84.22 13.36 5.424 86.90 6.787 86.00 9.065 84.17 13.50 5.447 86.89 6.823 85.91 9.129 84.11 13.65 5.470 86.86 6.859 85.86 9.194 84.06 13.79 5.493 86.86 6.895 85.91 9.260 84.00 13.94 5.516 86.86 6.932 85.93 9.326 83.95 14.09 5.540 86.83 6.970 85.92 9.394 83.90 14.25 5.564 86.80 7.007 85.91 9.462 83.83 14.
P MODEL PYROELECTRIC DETECTOR ABSORPTION VS. WAVELENGTH: Use this table for standard and slim profile painted (P) model pyroelectric detectors. Wavelength µm 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00 1.10 1.20 1.30 1.40 Absorption % 96.850 96.850 96.850 96.850 96.850 96.850 96.850 96.850 96.850 96.850 96.309 95.768 Wavelength µm 1.50 1.60 1.70 1.80 1.90 2.00 2.00 2.10 2.20 2.39 3.00 3.42 Absorption % 94.931 94.094 94.094 94.094 93.209 92.323 92.323 91.831 91.339 89.092 86.542 86.032 49 Wavelength µm 4.
HF MODEL PYROELECTRIC DETECTOR ABSORPTION VS. WAVELENGTH: Use this table for standard and slim profile high frequency (HF) model pyroelectric detectors. Wavelength Absorption µm % 0.200 58.05 0.210 57.59 0.220 58.33 0.230 57.76 0.240 57.18 0.250 56.55 0.260 56.08 0.270 55.63 0.280 55.35 0.290 55.07 0.300 54.79 0.310 54.56 0.320 54.21 0.330 53.91 0.340 53.74 0.350 53.03 0.360 52.95 0.370 52.62 0.375 52.42 0.380 52.21 0.390 51.81 0.400 51.25 0.410 50.94 0.420 50.77 0.425 50.65 0.430 50.52 0.440 50.36 0.
Wavelength Absorption µm % 2.067 60.27 2.070 60.28 2.073 60.24 2.077 60.24 2.080 60.24 2.083 60.25 2.087 60.26 2.09 60.25 2.094 60.25 2.097 60.25 2.100 60.23 2.104 60.24 2.107 60.26 2.111 60.24 2.114 60.22 2.117 60.22 2.121 60.22 2.124 60.23 2.128 60.23 2.131 60.22 2.135 60.21 2.138 60.20 2.142 60.20 2.145 60.17 2.149 60.16 2.153 60.16 2.156 60.17 2.160 60.18 2.163 60.17 2.167 60.15 2.171 60.13 2.174 60.12 2.178 60.13 2.182 60.16 2.185 60.16 2.189 60.15 2.193 60.14 2.196 60.12 2.200 60.12 2.204 60.11 2.
Wavelength Absorption µm % 3.120 59.47 3.128 59.47 3.135 59.47 3.143 59.48 3.151 59.47 3.158 59.47 3.166 59.47 3.174 59.48 3.182 59.48 3.189 59.48 3.197 59.47 3.205 59.47 3.213 59.47 3.221 59.46 3.229 59.47 3.237 59.48 3.245 59.47 3.253 59.47 3.262 59.46 3.270 59.46 3.278 59.46 3.286 59.46 3.295 59.46 3.303 59.46 3.312 59.46 3.320 59.46 3.328 59.46 3.337 59.45 3.346 59.45 3.354 59.44 3.363 59.43 3.372 59.41 3.381 59.41 3.389 59.40 3.398 59.39 3.407 59.37 3.416 59.36 3.425 59.37 3.434 59.38 3.443 59.39 3.
Wavelength Absorption Wavelength Absorption Wavelength Absorption Wavelength Absorption µm % µm % µm % µm % 6.365 57.97 7.870 57.74 10.31 60.33 14.93 43.14 6.396 57.83 7.918 57.73 10.39 60.59 15.11 42.72 6.428 57.83 7.966 57.73 10.47 60.88 15.28 41.67 6.460 57.76 8.015 57.73 10.56 61.18 15.47 40.14 6.492 57.75 8.065 57.73 10.65 61.51 15.65 38.41 6.525 57.77 8.116 57.73 10.73 61.88 15.84 36.53 6.557 57.69 8.167 57.74 10.82 62.26 16.04 34.41 6.591 57.67 8.218 57.74 10.91 62.62 16.24 32.05 6.624 57.73 8.
SCHEMATICS: 54