Specifications
24
[register number]_RCL_TARE[CR] REG XXX:[register contents][current
units][CR][LF]
Adds the specified register contents to
the TARE register.
RCL_XAVG[CR] [average weight] [current units]
AVX[CR][LF]. [2 second delay].
[number of samples taken]
###[CR][LF]. [2 second delay].
[standard deviation] [ current units]
SDEV[CR][LF]. [2 second delay].
[relative standard deviation]
COV[CR][LF].
Sends the results of a standard deviation
run. The TBAR mode, if in use, is
temporarily suspended while this
process is going on.
TYPE 4 COMMANDS (clock/calendar functions)
COMMAND RESPONSE PURPOSE
[six digit integer]
_ENTER_100_STORE[CR]
None
Sets the calendar function, if available,
to the specified date. The date must be
specified in the MMDDYY format and
no delimiters.
100_RCL[CR] REG 100: MM.DD.YY DATE[CR][LF] Returns the date.
[six digit integer]
_ENTER_101_STORE[CR]
None
Sets the clock function, if available, to
the specified time. The time must be
specified in the HHMMSS format using
military 24 hour time format.
101_RCL[CR] REG 101: [HH:MM:SS] TIME[CR][LF]
Returns the current time, using military
24 hour time format, once for each
request.
[1 to 6 digit integer]
_ENTER_89_STORE[CR]
None
Sets the auto send register to the
specified send time interval. For the
auto send feature to be operational, you
must set the auto send selection to ON
(see RS-232 Front Panel Configuration
section for details). The format used is
HHMMSS with no delimiters. For
instance, to set the auto send interval to
5 minutes and 25 seconds, you would
send this command string: 525 ENTER
89 STORE[CR]. Also you must have
the auto Send selection set to ON before
the auto Send will work (see RS-232
Front Panel Configuration Section).
89_RCL[CR] REG 089: [HH:MM:SS]
INTV[CR][LF]
Returns the auto SEND time interval
currently set.
[(1)(0)]_ENTER_79_STORE[CR] None A 1 in register 79 disables the date/time
stamp feature of any SEND. A 0 in
register 79, enables the date/time stamp.