22
[decimal number]_HI[CR] None
Sets the checkweighing high limit
register (register 88) to the decimal
number entered in the current units.
Register 88 is undefined at every
ON/OFF cycle.
[decimal number]_LO[CR] None
Sets the checkweighing low limit
register (register 87) to the decimal
number entered in the current units.
Register 87 is undefined at every
ON/OFF cycle.
[decimal number]_CONVERT[CR] None
Loads the number entered, as a "last
front panel units" relative multiplier,
into the CONVERT register and enters
the CONVERT mode. CLEAR exits the
convert mode and undefines the convert
register.
[decimal number]_CAL[CR] None
Loads a computed number based on the
entered number, as a "last front panel
units" relative multiplier, into the CAL
register and enters the CAL mode. The
display must be non-zero when this
command is issued. If the display is
zero, a 3 second "Err" message is
displayed and the balance returns to its
normal weighing mode. The current
weight reads as the number entered with
the appropriate sign. CLEAR exits the
CAL mode and undefines the CAL
register.
[integer]_PIECES[CR] None
Loads a computed number based on the
entered number, as a "last front panel
units" relative multiplier, into the
PIECES register and enters the PIECES
mode. The display must be non-zero
when this command is issued. If the
display is zero, a 3 second "Err"
message is displayed and the balance
returns to its normal weighing mode.
CLEAR exits the pieces mode and
undefines the pieces register.
[Decimal number]_ENTER_[integer]
_PIECES[CR]
None Accepts the decimal number as a weight
in "last front panel units" per integer
pieces. If the balance is not in PCS
mode, it goes into PCS mode using the
decimal number/the integer as the
weight. If used when the balance is
already in PCS mode, the display digits
are blanked for about 1 second while the
PCS an "last front panel units"
annunciators are lit. The balance then
resumes normal PCS operation using the
decimal number/the integer as the unit
weight.
[register number]_RCL_[(CONVERT)
(CAL) (PIECES)][CR]
REG XXX :[register contents][MULT.]
Puts the recalled register contents into
the specified mode register and enters
that mode.
TAVG[CR] None
Clears the TBAR registers and enters
the TBAR mode.
RESET_TAVG[CR] None
Clears the TBAR registers. If the
balance is in the TBAR mode, the
process keeps running.