Setup guide
5
4. New! Power up unit while keeping Unwrapped and Wrapped button pressed to enter
Statim Service Mode. This Service Mode is password protected, enter password to
continue (default password is: Unwrapped, Wrapped, R&P and Stop keys pressed in
this order).
Keypad functions at this time:
Unwrapped Key: Select next item in the menu.
Wrapped Key: Select previous item in the menu.
Rubber and Plastic Key: Enter current selection
Toggle using keypad through the menu selection to reach Calibration option and
press R&P key.
5. Confirm that the display appears similar to the example above.
The value in the lower right-hand corner of the display (CCC.C) represents Chamber
Temperature. The value in the upper left corner (PPP.P) is the Chamber Pressure. The
following two digits in the upper left corner of the display (PO) represent the Pressure
Sensor Offset value in hexadecimal followed by the theoretical chamber Pressure
calculated from Temperature (TTT.T). The two digits in the upper right corner of the
display (CO) represent the Chamber Offset value in hexadecimal.
NOTE: For software revisions higher than 4.15 (R415), in order to account for the
temperature drift of the pressure sensor, the field PO has two functions:
1. When chamber pressure is below 115kPa the offset will have one
value POA – Pressure Offset Atmospheric.
2. When chamber pressure is higher than 115kPa the PO field will
switch to indicate the POS – Pressure Offset Sterilization.
PPP.P PO TTT.T TO
CCC.C
Chamber
thermocouple
h
e
x
a
d
e
c
i
m
a
l
o
ffs
e
t
Chambertheoreticalpressure
[kPa]calculatedfrom
temperature
ChambertemperatureinºC
Pressure
offsetin
hexadecimal
Chamber
pressurein
kPa