Setup guide

11. Statim 2000S and 5000S Troubleshooting
Cycle Faults Software Revision R4xx, R5xx and R6xx
51
Cycle Fault #28
The cassette pressure rose above a ceiling value.
Pressure measurement failure. Gross pressure
and/or temperature sensor miscalibration. Unable to
depressurize cassette (blocked exhaust, failed valve,
etc.)
Cycle Fault #29
Not used
Cycle Fault #30
The cassette temperature failed to reach the
sterilization temperature within 15 seconds of the
theoretical cassette temperature calculated from the
measured cassette pressure reaching the
sterilization temperature.
Poor air removal during conditioning (exhaust
blockage).
Pressure and/or temperature sensor miscalibration.
Cycle Fault #50
For the Rubber and Plastics cycle, the chamber
temperature dropped below the sterilization
temperature, allowing for measurement error (i.e.
Tchm < 121°C).
Chamber temperature sensor miscalibration. Poor air
removal during conditioning (exhaust blockage). Not
able to generate steam or a leak in the system
Cycle Fault #51
For the Rubber and Plastics cycle, the chamber
temperature rose more than 4 C° above the
sterilization temperature, allowing for measurement
error (i.e. Tchm > 125°C).
Chamber temperature sensor miscalibration. Unable
to depressurize cassette (blocked exhaust, failed
solenoid valve).
Cycle Fault #52
For the Rubber and Plastics cycle, the theoretical
chamber temperature as calculated from the
measured chamber pressure was more than 2°C
below the measured chamber temperature, allowing
for measurement error.
Pressure and/or temperature sensor miscalibration.
Unable to depressurize cassette (blocked exhaust,
failed valve). Poor air removal (partial exhaust
blockage).
Cycle Fault #53
For the Rubber and Plastics cycle, the theoretical
chamber temperature as calculated from the
measured chamber pressure was more than 2°C
above the measured chamber temperature, allowing
for measurement error.
Pressure and/or temperature sensor miscalibration.
Unable to depressurize cassette (blocked exhaust,
failed valve). Poor air removal (partial exhaust
blockage).
Cycle Fault #54
For the Rubber and Plastics cycle, the theoretical
cassette temperature calculated from the measured
cassette pressure was below the sterilization
temperature, allowing for measurement error (i.e.
Pchm < 204.8 kPa).
Pressure and/or temperature sensor miscalibration.
Unable to generate steam, or there is a leak in the
system.