Setup guide
9. Statim 2000 Trouble-shooting
Cycle Faults – Software Revision R1xx, R2xx, R4xx, R5xx and R6xx
35
Cycle Fault #19
For software rev. R1xx, R2xx and R4xx the steam
generator calibration is invalid. For software rev.
R5xx and R6xx the validation thermocouple
calibration is invalid. In both cases a new steam
generator or validation thermocouple calibration is
required.
This occurs when a new controller board or
microprocessor has been installed. This may
also happen when the unit has been subjected
to a strong static discharge corrupting the
memory. Calibrate the boiler / validation
thermocouple
Cycle Fault #20
Software Revisions
R1xx or R2xx
The pump has failed to pump water into the steam
generator during a pre-vent pump time-out. The
steam generator temperature was greater than
140°C for 3.6 seconds after the pump was activated
to pump water to cool the steam generator.
This may be the result of a blocked steam
generator inlet tube, a failed or weak water
pump, a failed pump triac or steam generator
triac, faulty seal installation or failed solenoid
valve. Also may be the result of a defective
water quality sensor (insufficient water in the
reservoir)
Cycle Fault #21-#24
N/A
Cycle Fault #25
The software has failed to detect a need to pump
water within 90 seconds of the start of the cycle
The most probable cause is a blown thermal
fuse due to weak pump delivery and/or a
contaminated steam generator, or improperly
connected steam generator leads (loose or
unconnected). If the steam generator triac has
failed, the thermal fuse may have also failed.
Verify steam generator resistance (could be
failed open)
Cycle Fault #26
Software Revisions
R1xx, R2xx or R4xx
The sterilization phase has failed to start within 3
minutes of the cassette reaching the sterilization
temperature.
For software rev. R4xx CF26 is displayed when it
occurred in three consecutive cycles (Cycle
interrupted is displayed for the first two cycles).
Cycle Fault 26 counter is reset whenever a
successful cycle is completed.
This may be caused by improper boiler
thermocouple calibration, by a contaminated
steam generator, by a weak pump delivery or by
a faulty solenoid valve.
Also verify Vref to be 2.520V± 0.001V