User manual
28 • Here’s How! A QuickGuide for ColorQuartet
ColorQuartet 5.0 Pro
262
To focus-calibrate your flatbed scanner
1 • Place the supplied focus calibration target on the mounting plate.
2 • Open the Preview Setup window, choose Transmission, and make a preview scan of the
entire scan area.
3 • When the preview scan appears on screen, crop a small area (about 2 x 2 cm or 1" x 1") in
the part of the focus target with the parallel lines.
4 • Choose Calibrate Focus from the Calibration menu.
5 • Click on OK to confirm.
• ScanMate F10: Mount the target on the register pins.
• Other scanners: Tape the target across the base plate of the mounting plate,
lining it up parallel to the slit in the white tray under the mounting plate.
The calibration will automatically be carried out, and a message will
appear to tell you whether it was successfully completed.