User manual

28 • Here’s How! A QuickGuide for ColorQuartet
ColorQuartet 5.0 Pro
260
To white point-calibrate your drum scanner
1 • If you are calibrating a removable drum, carry out focus calibration first.
2 • Make a preview scan.
3 • When the Preview window appears, draw a crop frame that is about 10-20 mm (0.5"-1")
square, making sure that it includes only the clear drum or white paper.
4 • Choose Calibrate White from the Calibration menu.
5 • Click on OK to confirm.
This should also be done for the ScanMate 3000 the first time you cali-
brate it.
Transmission: scan a clear area of the drum.
Reflection: scan a piece of paper that is at least as white as the whitest part of
the originals you plan to scan.
The calibration will automatically be carried out, and a message will
appear to tell you whether it was successfully completed.