Instruction manual

3700R/3740 Refrigerated Sampler
Section 3 Programming Guidelines
3-10
Extended Programming – The extended programming mode
increases the number of available features, making more
complex sampling routines possible. For example, you can use
nonuniform time intervals, and create an intermittent sampling
schedule with the Stops and Resumes feature. Although all basic
programming features are available in the extended pro-
gramming mode, the less frequently used features have been sep-
arated into the extended mode to simplify the programming
process.
Although the basic and extended program sequences vary in
detail, you can divide both into four sections: Sample Pacing
(interval between samples), Sample Distribution (number of
samples in each bottle), Sample Volume (sample size in ml), and
Key Times (start times or, in the extended mode, stop and
resume times). These sections are noted in the Program
Sequence section of the diagram shown in Figure 3-6.