Operation Manual

Manual Indicator SA 250
pag. 6
PCG/C SA 250 Indicator Manual 23-03-07
Hi-Lo weighing continued
Setting the weight limits
HI
1. Press key . The symbol and the red LED at HI show up.
The present value of the upper limit is shown. The last figure is flashing
and may be changed now with the cursor key .
With key the next figure may be turned on and so on.
LO
2. Press key again. The yellow LED at Lo will light up.
The present value of the upper limit is shown.
Continue procedure as with the entering of the upper limit value.
3. By pressing key again, display reading will return to the weighing
mode.
A check of the present weight values only may be done by simply pressing
key three times.
This Lock-up function is permanently switched off at legal stamped scales.
The weight value of a weighing may be locked up temporarily.
This may be done in two different way’s: (setting in CAL menu):
F1 - The weight value will be locked up as soon as the reading comes to a
exact standstill, until the scale display returns to zero and comes to
a standstill at zero.
F2 - The weight value will be locked up as soon as a standstill within 1, 2 or 3
scale divisions is determined. However if the real weight value changes
for at least 1,2 or 3 scale divisions, the lock up will be set free .
On the condition that the Hold function has been selected in the Calibration
Menu, the function may be switched on and off by pressing key .
The display shows in the left corner above, which of the hold functions is in
operation with the symbols: F1 or F2.
WEEGTECHNIEK HOLLAND B.V. Website www.weegtechniek.nl Tel +31 36 522 20 30
Patroonsweg 23-27 Email info@ weegtechniek.nl Fax +31 36 522 20 60
3892 DA ZEEWOLDE The Netherlands
NET
NET
Hold mode (in weighing mode only, and the function has to be selected in CAL mode)
F 1
NET
F 2
NET