Lod . . . stays on display. At switching on the weight signal is too low.
Check the correct position of the weigh platform and possible damage of the
connecting cable and connector.
The loadcell could be damaged.
Execute a new one point adjustment procedure until the point 2 (zero setting).
The scale has been switched on, but zero setting is not possible since the
weight signal is higher or lower as 10% of the scale capacity.
Remove the load on the weigh platform. Check cable and connector.
Execute a new one point adjustment procedure until point 2 (zero setting).
The scale refuses zero setting.
The weight signal is outside the zero setting range(+ and - 2% scale capacity).
Tare function will operates correctly. Switch the scale off and on again so a
new zero (+ and - 10% scale capacity) may be settled.
Reading is gross
negative. Set zero and set tare not possible.
The scale has been switched on with a load on the platform and has set zero
with this load. Thereafter, the load has been removed and now the reading is
gross negative. Switch the scale off and on again with an empty platform.
The load on the platform is higher as the maximum scale capacity: overload.
Remove the load from the weigh platform. Check cable and connector.
The loadcell could be damaged.
The weight signal is far too negative. Check the correct position of the weigh
platform and clean the interior of the platform. Check for damage of the cable
and connector. The loadcell could be damaged.
Execute a new adjustment procedure.
Power supply excitation is too low, battery is empty.
Recharge the battery.
The Hi-Lo function cannot be selected with key F
.
- No weight values has been entered in the Hi-Lo mode.
Enter upper and down limit weight values with the key Hi-Lo.
- The Hi-Lo function is disabled in the USER mode.
The Lock up function cannot be selected with key F
.
- The Lock up function is disabled in the Calibration mode.
No data transmission or printing follows after operating the printer key.
- The transmission function is disabled in the USER mode
- Scale reading is at zero
- Scale reading has not been on zero with standstill after the last transmission.