Operation Manual

Manual Indicator SA 250
pag. 16
PCG/C SA 250 Indicator Manual 23-03-07
12. Add weights for the 3
e
point of adjustment upon the scale.
(This is not necessarily the max. weighing capacity)
13. Press key to enter the measured weight signal.
14. Enter the value of this weight with the keys
(increase) and (switch left)
The symbol F3 indicates this second point of adjustment.
15. Press key to enter the value of the weight
.
16. The display now asks for the value of the maximum scale capacity.
The display shows:
SP 2 , the figure at the right flashes and may be edited
now with key .
With key the next figure is made editable, and so on.
During normal operation of the scale the weight reading will switch off and
show “overload” at loadings above this weight value + 9 divisions.
17. Press key to enter the value of the maximum weighing capacity.
The display shows CAL during the entering procedure.
The adjustment procedure will now be finished and the scale will
automatically return to the normal weighing mode.
8. Remove the weight from the weigh platform.
19. Switch the scale off and on again. All keys will be accessible now.
The scale is ready to be used.