- Manual SATEX S V1.4 SA 235 - Waterproof weighing scale: IP 68 - Multi interval and high resolution of the reading user selectable Weegtechniek holland b.v. Patroonsweg 23 3892 DA Zeewolde The Netherlands Tel. Fax. Email Web +31 36 522 20 30 +31 36 522 20 60 info@satex.nl www.satex.
pag. 1 Manual SA 235 PCG/PC SA 235 Manual 01-03-2010 General Thank you for purchasing this waterproof SATEX precision weighing scale ! This scale is equipped with reliable technology and has user friendly operational software. You will use it many times and it will prove itself to be an easy, very speedy and accurate tool. If any questions raise that are not described in this manual, please address yourselves to your SATEX supplier, have a look at our website : www.satex.
pag. 2 Manual SA 235 Start up weighplatform (removable) 221 "On" switch Unpack the scale carefully Adjust the scale in level position with the 4 levelling feet and the level indicator in the front panel Connect to main supply, press the "On" switch for at least 1 second (with empty weigh platform ! !) 184 The scale sets zero automatically and now ready for use The built in battery allows for up to 170 hours continuously use without extern power supply.
pag.
pag. 4 Manual SA 235 Switching on lb kg Switch the scale on by pressing key for at least 1 second. The weighing platform has to be empty ! A "beep" sounds and the display shows all the Led's. g During the start up procedure the software version number is shown. lb kg g After that the date of the software appears. lb kg g Now the scale shall automatically set zero, as soon as there is an exact standstill of the weight signal. The Led's at : and at the weight unit light up.
pag. 5 Manual SA 235 Check weighing ( check weighing with 2 settable weight margins) Two weight margins can be entered: the down limit and the up limit. If the weight value is: - higher as the up limit, Led "HI" lights up - in between the up and down limit, Led "OK" lights up - lower as the down limit, Led "LO" lights up This weight control works for a gross, as well as for a tared (net) weight reading. In the user mode a "beep" sound may be connected to the weight margins.
pag. 6 Manual SA 235 Piece counting (With weight reading in kg's only: will be switched on automatically) The SA 235 features a simple piece counting program. Depending on the capacity of the scale, articles may be counted with a minimum weight of: If article weight is less, the counting will be mistaken ! It is advised with small articles to use a larger number of articles for the reference weighing. Scale capacity 1.5 kg 3 kg 7.
pag. 7 Manual SA 235 USER MODE The "user mode" can be entered by, with the scale switched on, pressing both the keys and together (start with key "F"). As soon as the display shows: PrE .. release both the keys Precision of the scale divisions In this menu section the accuracy of the scale divisions may be selected. Select with key one of the three available options: Single range One weighing range with 3000 scale divisions, e.g.
pag. 8 Manual SA 235 Automatic zero setting 0 . . .4 scale devisions (Auto zero, keeps weight reading automatically on zero) The zero value of the display reading is able to adjust itself automatically if small deviations from the zero value occur, like dust on the weigh platform, long-lasting loading or temperature changes. The auto zero will be active only within a selectable number of scale divisions: from 0 up to 4 divisions deviation from zero. Selection "0" switches off the auto zero function.
pag. 9 Manual SA 235 Display speed The speed of refreshments of the weightvalue in the display in case of rapid changes of the value, may be adjusted as: decreased or standard 0 = decreased, at stable weight the reading responds at a lower rate upon disturbances, is not practical for filling and dosing 1 = standard, the display follows increase and decrease of the weight complete and at high speed The setting is selectable with key Confirm the selection with key . , hereafter the user menu is closed.
pag. 10 Manual SA 235 ADJUSTING THE SCALE Select: 1 point or 3 point adjustment Whether the 1 point or 3 point adjustment procedure is in operation can be determined as follows: 1. switch the scale off if he is on 2. press together both the keys + and keep them pressed 3. switch the scale on with key and listen to the "beep" 4.1 if 1 x a "beep" sounds during switching on, the 1 point procedure is in operation (H2 in the menu) 4.2 if 1.
pag. 11 Manual SA 235 3 point adjustment: H1 returns from 5 to 0 returns from 5 to 0 returns from 5 to 0 Recommended for Mi and HRMi weight reading setting 1. the display shows H1 = 1 point adjustment procedure is selected. 2. now place accurate weights of 1/3 of the maximum weighing range on the weigh platform. (e.g. for a 3 kg scale: place 1 kg) 3. confirm with key , automatically followed by: 4. the weight adjustment will be executed, the most right display figure runs down from 5 to 0 . 5.