page 9
The “one key quick calibration function” requires the selection of a fixed weight
value, used for this quick calibration procedure.
Depending of the weighing capacity of the scale there is the following selecting
available, that will be shown in the display
Capacity Choice of
type of scale calibration weights
300 g 100, 200
, 300 g
600 g 200, 400, 500
, 600 g
1200 g 500, 1000
, 1200 g
1550 g 500, 1000
, 1500 g
3100 g 1000, 2000
, 3000 g
With key the available weight values may be shown one by one.
The figures in this table indicated as fat
will realise the best accuracy for
that scale, in relation to the full weighing range.
Make use of weights with sufficient accuracy, at least class M1.
With key the shown weight may be selected, after that the program will
automatically terminate the “menu program” : CAL.d appears.
Operate key : to return to the weighing mode.
Keep key : pressed for longer as 3 seconds.
start The display will show during a few seconds the value of the required,
accurate calibration weight, f.e.: 500 gram.
download weight Now the display asks for putting this weight upon the platform: dL. 500
After putting the weight in place, the display will flash during some seconds,
display flashes waiting for standstill and calculating the adjustment.
As the display stops flashing, weight adjustment has been done. The display
unload weight shows:
u
L , the calibration weight may be taken off the platform.
After taking off the weight, the display again flashes for some seconds
display flashes and the zero value will be adjusted.
The calibration procedure is done now : CAL.d , the scale automatically
completed switches back to the weighing mode.
dL = down Load
uL = un Load
CAL.d = Calibration done