PM296/RPM096 Power Quality Analyzers Installation and Operation Manual
LIMITED WARRANTY The manufacturer offers the customer an 24-month functional warranty on the instrument for faulty workmanship or parts from date of dispatch from the distributor. In all cases, this warranty is valid for 36 months from the date of production. This warranty is on a return to factory basis. The manufacturer does not accept liability for any damage caused by instrument malfunction.
shorting switches and is secured using an external strain relief to reduce mechanical strain on the screw terminals, if necessary. 7. Setup procedures must be performed only by qualified personnel familiar with the instrument and its associated electrical equipment. 8. DO NOT open the instrument under any circumstances. Read this manual thoroughly before connecting the meter to the current carrying circuits. During operation of the meter, hazardous voltages are present on input terminals.
Table of Contents Quick Start ...........................................................................................iv Chapter 1 Introduction .......................................................................1 1.1 1.2 About This Manual ..........................................................................1 About The PM296/RPM096 .............................................................1 Chapter 2 Installation .........................................................................8 2.
Quick Start TYPICAL INSTALLATION : Wiring Mode 4LL3, RS-485 Connection (see Sections 2.2.4 and 7.
SETUP (see Chapter 4 for full instructions) Setups can be performed directly on the front panel or via PComTest/PAS communication software - see Chapter 4 for details. menu bASc Performing Basic and Communications Setup Press CHG SELECT ENTER Press SELECT to activate middle window; press ▲▼ to scroll to option. Press SELECT to activate lower window; press ▲▼ to scroll to value. Press ENTER to save selected value.
Chapter 1 Introduction 1. 1.1 About This Manual This manual is intended for the user of the PM296/RPM096 Power Quality Analyzer. The PM296/RPM096 is a microprocessor-based instrument used for the measurement, monitoring, management and analysis of electrical parameters. This chapter gives an overview of this manual and an introduction to the PM296/RPM096. Chapter 2, Installation, provides instructions for mechanical and electrical installation.
Remote Display Module - Optional The RPM096 can be equipped with a Remote Display Module (RDM096) that provides local data display as well as setup capabilities. The RDM096 is connected to the RPM096 via a conventional DB15 connector (0.4/1.8/3.0 m cable supplied). The display is multi-page and is composed of three windows with high-brightness digital LEDs. One galvanically isolated direct current voltage input with options 300V, 100V or 20V.
Waveform Recorder for waveform capture and logging on different events. Along with the disturbance monitor it allows for capture and recording of various types of disturbances with a duration from one millisecond and up to tens seconds - transients, outages, sags, surges and deviations in voltage level.
User-selectable options are provided (see Section 4.11): 1) Power calculation mode (P.cAL) 2) Energy rollover value This option specifies the point at which the energy value rolls over to zero. 3) Thermal demand calculation option (thr.d) This option is used to enable or disable thermal power demand calculations.
Measured Parameters NOTE: Real-time values are measured over 1 cycle of fundamental frequency; Average values are of 8, 16 or 32 Real-time values Outputs Parameter Display Comm.
Parameter Display Comm.
Parameter RT Power Factor per phase RT Total Active Power RT Total Reactive Power RT Total Apparent Power RT Frequency RT Neutral Current RT Total Power Factor RT Auxiliary Current I4 3-phase RT Voltage & Current RT Voltage & Current unbalance RT DC Voltage Min/Max Logging Min/Max A, V, Frequency, total kW, kvar, kVA, Min/Max log for all real-time parameters Programmable Min/Max for harmonic values Voltage Disturbance Phase Rotation Phase Angles Day and Time Pulse Counters Remote Relay Control Inputs & Outp
Chapter 2 Installation 2. 2.
Figure 2-2 Mounting Brackets onto RPM096 Figure 2-3 RPM096 Mounting on DIN Rail Chapter 2 Installation 9
Figure 2-4 Mounting the RDM096 on Square/Round Cutout Figure 2-5 Connection of RDM096 to RPM096 10 Chapter 2 Installation
2.2 Electrical Installation Before installation ensure that all incoming power sources are shut OFF. Failure to observe this practice can result in serious or even fatal injury and damage to equipment. See General Schematic on page iv. 2.2.1 Power Source Connection The power source can be dedicated-fused, or from a monitored voltage if it is within the instrument’s power supply range. AC power supply: line to terminal 16; neutral to terminal 15.
Wiring Configuration Wiring (See parameter setup instructions in Section 4.
Figure 2-7 Four Wire WYE Direct Connection Using 3 CTs (3-element) Wiring Mode = 4LL3 or 4Ln3 Figure 2-8 Four Wire WYE Connection Using 3 PTs, 3 CTs (3-element) Wiring Mode = 4LL3 or 4Ln3 Chapter 2 Installation 13
Figure 2-9 Three Wire Open Delta Connection Using 2 PTs, 2 CTs (2-element) Wiring Mode = 3OP2 Figure 2-10 Four Wire Wye Connection Using 2 PTs, 3 CTs (2½-element) Wiring Mode = 3LL3 or 3Ln3 14 Chapter 2 Installation
Figure 2-11 Three Wire Open Delta Connection Using 2 PTs, 3 CTs (2½-element) Wiring Mode = 3OP3 Figure 2-12 Four Wire Delta Direct Connection Using 3 CTs (3 element) Wiring Mode = 4LL3 or 4Ln3 Chapter 2 Installation 15
2.2.5 Relays Six relays are provided for energy pulsing, alarms or remote control. Figure 2-13 Relays Connection 2.2.6 Status Inputs 12 optically isolated status inputs are provided for status monitoring and external synchronization of power demand period and time.
2.2.7 Analog Output The PM296/RPM096 provides two optically isolated analog outputs with an internal power supply and current output options of 0-20 mA and 4-20 mA (current loop load of up to 500 Ohm), 0-1 mA and ±1 mA (current loop load of up to10 kOhm). Figure 2-15 Analog Output Connection 2.2.
Chapter 3 Using The Menus 3. Press and release SELECT to enter the setup mode. The primary menus will appear: StA OPS CHG - Status Information Menu (see Chapter 6) - Setup Options Menu - Setup Change Menu (see Chapter 4) Press SELECT again to activate the window of the desired primary menu. Press ENTER . Select CHG to initialize or modify the instrument setup, or to clear the accumulated values stored in the instrument. Entry to this menu can be protected by a password.
Figure 3-1 Menu Structure SELECT ENTER Status Information StA Setup Options ENTER OPS Setup Change Password CHG PASS Reset Functions rSt Real Time Clock Phase Rotation PHAS Phase Angles rtc AnGL Basic Setup Relay Status bASc rEL Port Setup Status Inputs Port Digital Inputs dinP Counters Cnt (if ordered) Analog Outputs (if ordered) Analog Expander Aout St.In Counter #1 Cnt.1 ... Counter #16 Cn.
Chapter 4 Setup Menus 4. CHAPTER 3 SETUP MENUS NOTE: Instrument setup can be performed directly on the front panel using the setup menus or via communications using PComTest or PAS communication software, supplied with your instrument. Alarm/Event Setpoints can be programmed only through communications. For information on using PComTest or PAS, refer to the user documentation provided.
Table 4-1 Basic Setup Options (∗ default setting) Code ConF Parameter Wiring mode Options 3OP2 4Ln3∗ 3dir2 4LL3 3OP3 3Ln3 3LL3 Pt Ct Ct.Au dc.OF dc.FS d.P n.dp A.dP t.con buF rSt PrE.C PT ratio 1.0∗ 6,500.
Code rEc.C Parameter Options The number of 0 to 2560 cycles in a waveform (0*) series Freq LoAd Nominal frequency Maximum demand load current Description The total number of waveform cycles to be recorded on either event occurrence. Will be rounded to a nearest bigger number multiple of 16. 0 = auto-select The nominal line frequency The maximum demand load current used in TDD calculations (0 = CT primary current) 50, 60 Hz** 0-10,000A (0∗) ** 60 Hz default for North America; elsewhere, default is 50Hz.
4.2 Communications Port Setup Menus SELECT CHG ENTER Prt.1 ENTER SELECT CHG ENTER Prt.2 ENTER These menus allow you to access the communications port options that the PM296/RPM096 uses to communicate with a master computer or a printer. Table 4-2 lists the communications options, their code names and applicable choices. From the main menu, select the menu for the port you want to configure.
Code dAtA Parameter Data format H.Sh Incoming flow control (handshaking) Options 7E 8n∗ 8E nonE∗ SOFt Hard Ctrl Outgoing flow control (RTS/DTR) nonE∗ dtr rtS Printout period 1∗, 2, 5, 10, 15, 20, 30, 60 min Code Prt.2 Parameter Communications protocol rS Interface standard Addr Address bAud Baud rate dAtA Data format CPtb ASCII compatibility mode Options ASCII∗ rtu dnP3 422 485∗ 0∗-99 ASCII 1∗-247 Modbus 110 300 600 1200 2400 4800 9600∗ 19.20 38.40 7E 8n∗ 8E diS∗, En Prn.
4.3 Digital Inputs Setup Menu SELECT CHG ENTER dinP ENTER This menu is used to set up the 12 digital inputs provided by the PM296/RPM096. Each digital input can be allocated as: - a status input to monitor external contact status, or - a pulse input to sense pulses provided by an external source. One of these can be configured to receive an external synchronization pulse indicating the beginning of a new demand interval for power demand measurements.
4.4 Pulse Counters Setup Menu SELECT CHG ENTER Cnt ENTER This menu is used to configure the 16 pulse counters provided by your PM296/RPM096 instrument. Any counter can be connected to one of the 12 digital inputs, to count incoming pulses (in this event the connected digital input must be allocated as a pulse input as directed in Section 4.3) or to count a wide variety of events via setpoints. Each counter can be independently scaled (weighted) by specifying a scale factor in the range of 1 to 9999.
4.5 Analog Output Setup Menu [This section is relevant to instruments ordered with this option] CHG SELECT Aout ENTER ENTER This menu allows you to set up an output value and its zero and full scales for either of the two internal analog output channels. Table 4-3 explains the analog output setup options, and Table 4-4 lists all measurement parameters that can be directed to analog output. Output parameter Zero-scale output An. 1 Outp rt U1 An. 1 Lo 0 Full-scale output An.
parameter. For signed (bi-directional) values, such as powers and signed power factor, the scale is always symmetrical with regard to 0 mA, and the full scale corresponds to +1 mA output for positive readings and to -1 mA output for negative readings. For these, the zero scale (0 mA output) is permanently set in the instrument to zero for all parameters except the signed power factor for which it is set to 1.000, and may not change.
Code rt P rt q rt S rt PF rt PF.LG rt PF.Ld rt Au.C rt nEU.C rt Fr rt U.dC Ar U 1 Ar U 2 Ar U 3 Ar U.AG Ar U.L12 Ar U.L23 Ar U.L31 Ar U.L.AG Ar C1 Ar C2 Ar C3 Ar C.AG Ar P Ar q Ar S Ar PF Ar PF.LG Ar PF.Ld Ar Au.C Ar neU.C Ar Fr Acd.P.i Acd.P.E Acd.q.i Acd.q.E Acd.
Wiring via PTs (PT Ratio > 1): Vmax (690 V input option) = 144 × PT Ratio [V] Vmax (120 V input option) = 144 × PT Ratio [V] Pmax = (Imax × Vmax × 3)/1000 [MW x 0.001] @ wiring modes 4Ln3, 3Ln3 Pmax = (Imax × Vmax × 2)/1000 [MW x 0.001] @ wiring modes 4LL3, 3OP2, 3dir2, 3OP3, 3LL3 NOTE: Pmax is rounded to nearest whole kW units. The actual frequency range is 45.00 - 65.
4.7 Pulsing Output Setup Menu SELECT CHG ENTER PulS ENTER This menu allows you to program either of the six relays provided by your PM296/RPM096 instrument to output energy pulses. Available pulsing parameters are listed in Table 4-5. To select a pulse relay: rEL.1 Ac.Ei 1 Use the up/down arrow keys to scroll to the desired relay. The pulsing parameter assigned to the relay is displayed in the middle window, and the amount of unit-hours per pulse is displayed in the lower window.
4.8 Timers Setup Menu SELECT CHG ENTER t-r ENTER This menu allows you to access the four interval timers provided by the PM296/RPM096 which can trigger setpoints on a user-defined time interval basis. This is useful for continuous data logging at specified time intervals in order to produce trend and load profile graphs. Each timer has a time interval range up to 9999 seconds at a one-second resolution and runs independently. The timer accuracy is about ±0.05 sec.
4.9 RTC Setup Menu SELECT CHG rtc ENTER ENTER This menu allows you to view and set the time, date and day of week in the onboard Real Time Clock (RTC), and to modify the Daylight Savings Time (DST) settings for your time zone. hour 11.52 45 dAtE 25.07. 99 dAY The time is displayed as HH.MM.SS, where the hour and minute are shown in the middle window separated by a dot, and the seconds - in the lower window. The date is displayed as per the user definition (YY.MM.DD, MM.DD.YY, or DD.MM.
To select an option sub-menu: From the upper window, use the up/down arrow keys to scroll to the desired sub-menu (time, date, weekday, or DST). To change time, date, day of week (not seconds) or DST setting: Press SELECT to activate the desired item. When in the time setup sub-menu, the hour and minutes indications are now frozen to allow you to adjust them. Use the up/down arrow keys to set the value. Set the other items in the same manner.
4.10 Display Setup Menu SELECT CHG diSP ENTER ENTER This menu allows you to view and change display properties. Table 4-6 lists available options with their code names and applicable ranges. Table 4-6 Display Options (∗ default setting) Display diSP UPdt 0.5 diSP ScrL 5 Code Parameter Options UPdt Display update time 0.1 - 10.0 s (0.
To select a display option: Press SELECT to activate the middle window, and then use the up/down arrow keys to scroll to the desired option. To change the display option: Press SELECT to activate the lower window. Use the up/down arrow keys to set the desired option. Press ENTER to store your new setting or press ESC to leave your previous setting unchanged. To quit the display setup menu: From the middle window, press ESC or ENTER . 4.
Code Parameter An.Ou Analog output option (see Section 4.5) An.EP Analog expander option (see Section 4.6) bAtt Battery mode Options Description nonE∗ 0-20 4-20 0-1 -1-1 nonE∗ 0-20 4-20 0-1 -1-1 OFF∗, On No analog output 0-20 mA 4-20 mA 0-1 mA ±1 mA No analog expander (output disabled) 0-20mA 4-20mA 0-1mA ±1 mA Switches the backup battery OFF/ON Power calculation mode (P.
4.12 Access Control Menu SELECT CHG AccS ENTER ENTER This menu can be only accessed via the Setup Change Menu (CHG). It is used in order to: • • • change the user password enable or disable password check from the front panel keypad enable or disable password protection for downloading setups and resetting data through communications To view an option setting: Press SELECT to activate the middle window. Use the up/down arrow keys to scroll to the desired option (PASS, CtrL, Port).
4.13 Reset Menu CHG SELECT ENTER rSt ENTER This menu allows you to reset to zero the accumulators and Min/Max registers in your instrument. The menu can be only accessed via the Setup Change Menu (CHG). If the reset is disabled from the Basic Setup Menu (see Section 4.1), you will not be able to enter this menu. The following designations are used in the menu to specify a data location to be reset: A.dnd P.dnd dnd EnrG Lo.Hi tOU.E tOU.d Cnt Cnt.1 Cn.
Chapter 5 Data Display 5. 5.1 Navigating in the Display Mode The front panel has a simple interface that allows you to display numerous measurement parameters. For easier reading, the parameters are divided into groups, each accessible by a designated key. These are: PM296 Common measurements Max demand measurements Min/Max measurements Total Harmonic measurements Individual Harmonic measurements Energy measurements - Page 1. Main screen (no selection key) Page 2. Main screen (no selection key) Page 4.
The display is updated approximately twice per second; you can adjust the display update rate via the Display Setup Menu (see Section 4.10). Tables 5-1 and 5-2 list all displayed parameters and their LED indicators for the PM296 front panel display and RDM096 remote display module. Auto Scroll If display Auto Scroll option is enabled (see Section 4.10), the common measurements display (main screen) will scroll automatically after 30 seconds of uninterrupted use.
Selecting Max Demand Measurements In the RDM096, press the MIN/MAX key. Use the up/down arrow keys to scroll through Max demand measurements. Selecting Min/Max Measurements Press the MIN/MAX key. Use the up/down arrow keys to scroll through Min/Max measurements. Selecting Total Harmonic Measurements In the RDM096, press the H/ESC key until the THD/TDD LED is illuminated. Use the up/down arrow keys to scroll through harmonic measurements.
Fast Reset of Accumulated Data Select a display page where the data you want to reset is displayed. To reset: • Min/Max log registers: select a Min/Max page from the Min/Max measurements display (where Lo or Hi is displayed in the PAGE window in the PM296 or at the left in the lower window in the RDM096).
44 Page Window 1 1 1 2 2 2 2 2 3 3 3 3 3 3 3 4 4 4 4 4 4 4 5 5 5 5 5 5 5 9 10 11 11 7 1 2 3 7 1 4 8 9 10 11 7 2 5 8 9 10 11 7 3 6 8 9 10 11 6 6 6 6 6 6 6 6 11 1 2 3 4 5 6 8 6 9 6 10 7 7 7 11 1 2 PAGE LEDs Parameter Digits Total kvar Total kVA Total power factor 5 5 4 PAG.1 L-L Voltage L12 Voltage L23 Voltage L31 Ph.L1 Voltage L1/L12 Current L1 kW L1 kvar L1 kVA L1 Power factor L1 Ph.L2 Voltage L2/L23 Current L2 kW L1 kvar L2 kVA L2 Power factor L2 Ph.
Page Window 7 7 7 7 7 7 7 8 8 8 8 8 9 9 9 9 9 10 10 10 10 11 11 11 11 11 11 11 11 11 11 3 4 5 6 7 8 10 11 7 4 5 6 11 7 4 5 6 7 4 8 5 10 1 2 3 4 5 6 8 9 11 1 1 1 1 1 1 1 1 1 1 2 2 2 11 1 2 3 4 5 6 7 8 9 10 11 1 2 3 Chapter 5 Data Display PAGE LEDs Parameter Digits Unit Voltage THD L3 Current THD L1 Current THD L2 Current THD L3 Neutral current DC voltage Auxiliary current I4 4 4 4 4 4 5 5 tdd Current TDD L1 Current TDD L2 Current TDD L3 4 4 4 % % % % A V A/mA Label Label % % % Label Label PAG
46 Page Window 2 2 2 2 2 2 2 4 5 6 7 8 9 10 1 1 1 1 1 1 1 11 7 10 1 2 3 4 5 6 20 20 20 20 20 20 20 7 10 1 2 3 4 5 6 1 1 1 1 1 1 1 1 1 1 7 10 1 2 3 4 5 6 8 9 11 20 20 20 20 20 20 20 7 10 1 2 3 4 5 6 PAGE LEDs Parameter Digits Maximum current L1 4 Maximum current L2 4 Maximum current L3 4 Maximum frequency 4 Maximum total kW 5 Maximum total kvar 5 Maximum total kVA 5 Individual Odd Voltage and Current Harmonics H03-H39 PAG.
Page Window 20 20 20 8 9 11 1 1 1 1 1 2 2 2 2 3 3 3 3 11 1,4,7 2 5,8 3 6,9 2 5,8 3 6,9 2 5,8 3 6,9 1-16 1-16 1-16 10 11 2,5,8 PAGE LEDs Parameter Digits Harmonic H39 total kW Harmonic H39 total kvar Harmonic H39 total power factor Total Energies 5 5 4 MVAh IP. MWh import IP. Mvarh import EP. MWh export EP. Mvarh export U-h Volt-hours A-h Ampere-hours TOU Energy Registers trF.1 - tF.16 rEG.1 - rG.16 Tariff register reading 9 PAG.
The RDM096 Data Display The display windows are labeled in the table as follows: 1 = upper window, 2 = middle window, 3 = lower window.
Page 17 17 17 18 18 18 19 19 19 Window 1 2 3 1 2 3 1 2 3 9 9 9 10 10 10 11 3 1 2 3 1 2 3 1 11 2 11 3 1 1 1 2 2 2 3 3 3 4 4 4 5 5 5 6 6 6 7 7 7 3 1 2 3 1 2 3 1 2 3 1 2 3 3 1 2 3 1 2 3 1 2 3 Chapter 5 Data Display Arrow LED Parameter Digits 01H (Fundamental harmonic) PF H01 total power factor 4 kW H01 total kW 4 V1/V1-2 Fund. Harmonic voltage L1/L12 4 V2/V2-3 Fund. Harmonic voltage L2/L31 4 V3/V3-1 1H Fund. Harmonic voltage L3 4 A1 Fund. Harmonic current L1 4 A2 Fund.
Page 8 8 8 50 Window 1 2 3 1 1 1 2 2 2 3 3 3 4 4 4 1 2 3 1 2 3 1 2 3 1 2 3 1 1 1 1 2 3 20 20 20 1 2 3 1 1 1 1 2 3 20 20 20 1 2 3 1 1 1 1 2 3 20 20 20 1 2 3 1 1 1 2 Arrow LED Parameter Digits A NEUT Maximum neutral current 4 Hz Maximum frequency 4 kvar Maximum total kvar 4 Total Harmonic Measurements THD/TDD V1/V1-2 Voltage THD L1/L12 4 V2/V2-3 Voltage THD L2/L23 4 V3/V3-1 thd. Voltage THD L3 4 A1 Current THD L1 4 A2 Current THD L2 4 A3 thd.
Page 1 2 2 2 3 3 4 4 4 5 5 5 6 6 7 7 1-16 1-16 1-16 Window 3 1 2 3 1 3 1 2 3 1 2 3 1 3 1 3 1 2 3 Arrow LED Mvarh MVAh MWh Mvarh MWh/ Mvarh/ MVAh Parameter MWh import rE.En. IP. Mvarh import AP.En. MVAh Ac.En. EP. MWh export rE.En. EP. Mvarh export U-h Volt-hours A-h Ampere-hours TOU Energy Registers rEG.1 - rG.16 trF.1 - tF.
5.3 Self-Test Diagnostics Display The PM296/RPM096 periodically performs self-test diagnostics during operation. If the instrument fails the test, it discards the last measurement results, and an error code is displayed for one second on all LEDs. Error codes are listed in Table 5-2. Frequent failures may be the result of excessive electrical noise in the region of the instrument. If the instrument resets itself continuously, contact your local distributor.
Chapter 6 Viewing Status Information 6. Through the Status Information Menu (STA), it is possible to view the status of various instrument features. 6.1 The Status Information Menu SELECT StA ENTER To enter the Status Information Menu: From the display mode, press SELECT to enter the Primary Selection Menu. Press SELECT to activate the StA window. Press ENTER. To select a display page: Press the up/down arrow keys to scroll through the display pages.
6.2 Status Display Formats Tables 6-1 and 6-2 list all the displays available from the Status Information Menu. The PM296 Status Display The display windows are labeled in the table in the direction up-to-down and leftto-right. Table 6-1 Status Information Display for the PM296 Page 1 1 1 2 2 2 2 3 3 3 4 4 4 7 7 Window 4 5 6 4 1 2 3 1 2,5 4,6 1 3,6,9 3,6,9 1 3 22 22 23 23 1 3 1 3 Parameter PHAS. rOt. Phase rotation (POS/NEG/ERR) AnGL Phase L1/L12 angle Phase L2/L23 angle Phase L3/L31 angle rEL. 1.2.3.4.
Page 3 3 3 4 4 4 5 5 5 6 6 6 7 7 Window 1 2 3 1 2 3 1 2 3 1 2 3 1 3 22 22 23 23 1 3 1 3 Parameter rEL. 5.6. Relay #5-6 status St.In 1.2.3.4 Status inputs #1-#4 St.In 5.6.7.8 Status inputs #5-#8 St.In 9.a.b.C Status inputs #9-#12 Cnt.1 Counter #1 ... Cn.
Chapter 7 Communications 7. CHAPTER 6 USING A PRINTER A full description of the communications software is found in the PM296/RPM096 ASCII, Modbus and DNP 3.0 Communications Manuals provided on electronic media. 7.1 Using a Printer 7.1.1 Configuring the Port for Printer Set the printer mode and desired printout period in the communications port setup (see Section 4.2). The baud rate and data format should be configured as those on the printer.
Table 7-1 Printout Record Parameters Line 1 1 1 1 1 1 1 1 1 1 2 Place 1 2 3 4 5 6 7 8 9 10 1 Heading V1/kV1 V2/kV2 V2/kV3 A1 A2 A3 kW/MW kvar/Mvar kVA/MVA PF THDU1 Parameter Voltage L1/L12 Voltage L2/L23 Voltage L3/L31 Current L1 Current L2 Current L3 Total kW Total kvar Total kVA Total power factor Voltage THD L1/L12 2 2 2 2 2 2 2 2 2 3 3 3 3 2 3 4 5 6 7 8 9 10 1 2 3 4 3 5 3 3 6 7 THDU2 THDU3 THDI1 THDI2 THDI3 Hz A_NEU U_UNB I_UNB A1_MD A2_MD A3_MD kW_MD/ MW_MD kVA_MD/ MVA_MD +MWh -MWh Voltage T
7.2 Connections Figure 7-1 RS-232 and RS-422/RS-485 Terminal Blocks RS-232 - up to 15 meters distance, 1 PC/PLC/Printer to 1 PM296/RPM096, by flat 2 or twisted pair cable of 0.
Computer Connections for RS-232: COM1 Figure 7-8 RS-232 Simple 3-Wire Connection, 25-pin Figure 7-9 RS-232 Simple 3-Wire Connection, 9-pin Figure 7-10 RS-232 Hardware Handshaking Connection, 25-pin Figure 7-11 RS-232 Hardware Handshaking Connection, 9-pin Chapter 7 Communications 59
Computer Connections for RS-422: COM2 Computer Connections for RS-485: COM1 Figure 7-12 RS-422/RS-485 Connections Using Converter 60 Chapter 7 Communications
Computer Multidrop Connections for RS-485: COM1 Figure 7-13 RS-485 Multidrop Connection, 25-Pin Connector Figure 7-14 RS-485 Multidrop Connection, 9-Pin Connector Chapter 7 Communications 61
Computer Multidrop Connections for RS-422/RS-485: COM2 Figure 7-15 RS-422 Multidrop Connection, 25-Pin PC Port Figure 7-16 RS-422 Multidrop Connection, 9-Pin PC Port Figure 7-17 RS-485 Multidrop Connection, 25-Pin PC Port Figure 7-18 RS-485 Multidrop Connection, 9-Pin PC Port NOTE: Where an RS-232/RS-422 converter is used, R1 is not applicable since it is built in to the converter.
Appendix: Technical Specifications 8. ions Input and Output Ratings CATEGORY II POLLUTION DEGREE 2 (IEC 664) 3 galvanically isolated current inputs CATEGORY II POLLUTION DEGREE 120 V AC INPUT USING PT (up to 120+20% V line-to-line voltage) Burden: < 0.15 VA 690 V AC DIRECT INPUT (up to 690+15% V line-to-line voltage or up to 500 V line-to-neutral voltage) Burden: < 0.35 VA INPUT USING PT (up to 120+20% V line-to-line voltage) Burden: < 0.03 VA 1A INPUT via CT with 1 A secondary output Burden: < 0.
Input and Output Ratings Current input terminal Rated 10A 250V Pitch 10 mm Wire 3 mm2 (10 AWG) Voltage input connector Rated 32A 690V Pitch 9.5 mm 2 Wire max. 4 mm (10 AWG) Communications Two optically isolated serial ports: COM1 RS-232 (9-pin D-type connector) (COM1: only one line operable at a time) COM1 RS-485 (3-pin connector) COM2 RS-422/RS-485 (5-pin connector) Standard 5 mm pitch (UL recognized 7463) SCREW M3 Maximum wire diameter 2.
Storage temperature -25°C to 80°C (-13°F to 176°F) Humidity 0 to 95% non-condensing Construction Instrument body CASE ENCLOSURE: plastic ABS/PC Blend (UL recognized UL94V0) FRONT PANEL: plastic PC -Film (UL recognized UL94V0) Instrument weight 2.65 kg (6 lbs.
Measurement Specifications Parameter Voltage Line current Full scale Accuracy, % @ input Rdg FS Conditions 120VxPT For L-N reading 0.1 0.11 10% to 120% FS @ 120V and 3OP2/3OP3 400VxPT wiring modes @ 690V 208VxPT For L-L reading @ 120V except 690VxPT 3OP2/3OP3 @ 690V wiring modes CT PRIMARY CURRENT 0.2 0.007 1% to 200% FS Range 0 to 999,000 V Wiring via PTs (PT>1): 0.001 kV @ 0.001kV to 9.999 kV ≤0.1% @ 10.00 kV to 999.0 kV Starting voltage 1.5% FS 0 to 9999 A Active power 0.36×PT×CT @ 120V |PF| ≥ 0.
Parameter current Auxiliary current DC Voltage Full scale @ input Rdg Accuracy, % FS Conditions Range AUXILIARY CT PRIMARY 0.1 CURRENT 0.1 1% to 200% FS 0 to 9999 A/mA 20, 100, 300 VDC (upon order). Can be scaled up to 9999. 0.3 1% to 100% FS 0 to 9999 VDC Ampere demand kW demand (block & sliding) kvar demand KVA demand (block & sliding ) Total Harmonic 999.
Parameter Full scale @ input Rdg Accuracy, % FS Conditions Range Volt-hours 0 to 999,999.999 kVh Ampere-hours 0 to 999,999.999 kAh Display resolution (%Rdg) @ range 1 kVAh @ 1 to 999,999,999 kVAh RPM096: 1 kVAh @ 1 to 999,999 kVAh 10 kVAh @ 1000 to 9,999.99 MVAh 100 kVAh @ 10,000 to 99,999.9 MVAh 1MVAh @ 100,000 to 999,999 MVAh PM296: 1 Vh @ 1 to 999,999,999 Vh RPM096: 1 Vh @ 1 to 999,999 Vh 10 Vh @ 1000 to 9,999.99 kVh 100 Vh @ 10,000 to 99,999.
Index A accuracy, 32, 69 active energy, 67 active power, 36, 37, 66 ampere demand, v, 21, 31, 44, 49, 57 analog output, i, 2, 22, 27, 28, 30, 37 B basic setup, 20 C communications, 69 connections, 1, 56 CT, v, 1, 11, 21, 22, 29, 30, 35, 41, 63, 66, 67, 68 current inputs, 63 current transformers, 69 D demand, v, 2, 6, 21, 22, 25, 29, 30, 31, 39, 43, 44, 49, 57, 67 demand period, v, 21 diagnostic, 7 digital inputs, 3, 25, 26 display, 1, 2, 35, 36, 39, 40, 41, 43, 47, 48, 51, 53, 54, 64 display mode, 35, 39