Specifications

4.2. Internal Calibration - isoCAL
ISO
CAL
ISO
CAL
ISO
CAL
Target
Drift in the readout caused by
fluctuations in the ambient conditions,
such as temperature
^ User Problem
A balance needs to be calibrated
2
and
adjusted.
- on regular base.
- after temperature change.
- every time a balance was physically
moved.
- every time it was leveled.
Very often the user is obliged to
keep a calibration history.
b
!
Sartorius Solution
Secura is equipped with isoCAL. IsoCAL is
an internal calibration and adjustment per-
formed automatically with the fluctuation
of temperature or by a timed interval.
The internal calibration is performed by
a motorized weight located inside the
balance. Every time isoCAL is performed, a
calibration report is generated and stored
for traceability. Retrieval of this calibration
report is called Cal Audit Trail. External
Calibrations are also stored in this memory.
User Benefits:
Better
ily
:
dri its own Within seconds
messag'ein the display whentalahee
"'rib measurement-ean be compromised
influenced Ijke air flow, dust and finger-prints, ...
ca.lfb'rations'(D.epending .on cu'stortiers
seconds callBrated and adjusted.- - .
automatically.
-
;
L)s%frgets.rftlg9Sed;fcQ"rri stress to monitor proper calibration, i-~ • ' \
Esteem-1[\%f. \-_ - ' V" ._ '
'•hs6e4t
4
JSpiafl0f*>twithe art weighing technology.
How to Demonstrate
10 sec. after leveling, Secura will automatically self-calibrate (isoCAL).
- Safety level is set to "High" and
- isoCAL is set to "Automatic" {/* -* Calibration | Adjustment -» isoCAL).
- Within this 10 sec. "isoCAL" is shown. Touching this will also start the isoCAL.
- Behind the menu button ^^g you will find "CAL" on the left side. Behind that you
will find "CAL-lntern" which also starts the intern calibration and adjustment routine.
After the routine has ended a calibration report will appear. See chapter 9 "Customers
section" for example to show.
10
Calibration history is stored sequentially (dated) in Secura in day by day order.
{/" -* Calibration | Adjustment -» Calibration report). User will find up to 99 reports
per day stored for 30 days.
2
The commonly used term ..Calibration" means actually calibration and adjustment. While the meaning
of "calibration" is actually describing the determination of a deviation, "adjustment" is describing the
correction of the determined deviation. In this chapter we will use both terms in its original meaning.
3
Intensity is depending on the settings for "Safety level".