Operation Manual

5
1. Set the volume by rotating the operating
button (clockwise to decrease the volume
and counterclockwise to increase).
NOTE: Do not rotate the operating button
outside the volume range of the pipette.
7.2. Sealing and ejecting tips
To ensure the maximum accuracy and precision,
Sartorius tips are recommended to be used with
Proline®Plus pipettes. Before fitting a tip ensure
that the pipette tip cone is clean. Press the tip on
the tip cone of the pipette.
Eject the tip by pressing the tip ejector with your
thumb. Make sure that the tip is disposed into a
suitable waste container.
7.3. Protective tip cone filters
The tip cone design of the Sartorius pipettes (>10
µl) allows the use of Safe-Cone Filters in the tip
cones. These removable filters prevent liquids
and aerosols from entering the pipette and thus,
protects the pipette from contamination and
damage.
The Safe-Cone Filters are available as Standard
or Plus versions. It is recommended to use the
Standard filter for general applications and the
Plus filter for more demanding applications such
as cell culture, bacterial and virological work and
molecular biology. Filters need to be exchanged
regularly. The interval for filter exchanging is
application dependant but the recommendation is
to change the filter daily (after 50 - 250 pipetting
cycles) and always in case of over-aspiration. (See
Chapter 2 for ordering information.) The Filter
Foreceps should be used to avoid touching the
soiled filters by hand. See Chapter 2 for ordering
information. Make sure that the filter is disposed
into a suitable waste container. Clean the tip cone
if needed and put the new filter in place.
8. Pipetting techniques
Two basic pipetting techniques, forward and
reverse pipetting, are associated with Proline®Plus
pipettes. Forward pipetting is the most common
used pipetting technique. The technique employs
Tip ejector
+