User manual
84
USB Port
User Manual for Secura, Quintix, Practum
Special Outputs
Position 1 2345678910111213141516
******––******CRLF
or ******Low*****CRLF
or ******High****CRLF
* Space Cal.Ext. External adjustment
– – Final readout Cal.Int. Internal adjustment
Low Underweight CR Carriage return
High Overload LF Line feed
Error Messages
Position 12345678910111213141516
***Err**/###****CRLF
* Space ### Error code number
Example: Output of Weight Value + 111.255 g
Position 1 2 3 45678910111213141516
* * * 111. 255* g* * CRLF
+* * 123. 5[ 6]
1)
g**CRLF
Position 1: Plus +, minus –, or space
Position 2 – 10: Space or weight value with decimal point;
leading zeros are output as spaces.
Position 11: Space
Position 12 – 14: Characters for unit of measure or space
Position 15: Carriage return
Position 16: Line feed
1)
Settings “PC - SBI” and “USB RS232 SBI” for ID of non-verified digits:
In the “SBI” setting, the non-verified display digit is not automatically identified.
Please take the corresponding measures or adjust the settings on the peripheral
device.