User manual
54
Density
User Manual for Secura, Quintix, Practum
t To save the weight value, select v.
t Remove the sample from the weighing pan and place it in the sample holder of
the density determination kit using forceps.
! Ensure that the sample is completely submerged in the buoyancy liquid and
that no air bubbles have formed on the sample.
Max 220 g d
=
0.0001 g
4.4787
Weight in medium
END
+
Net
y The weighing display shows the step Weight in medium.
t To save the weight value, select v.
Max 220 g d
=
0.0001 g
11.86
END
+
y The balance calculates the density of the sample and displays this value.
t To exit the function, select x.
y The application switches back to its original state. If the GLP printout function
is activated, the GLP printout is automatically printed (see “Printout” on
page31).
There is an online animation for this application. Click here to view it: