User manual

54
Density
User Manual for Secura, Quintix, Practum
t To save the weight value, select v.
t Remove the sample from the weighing pan and place it in the sample holder of
the density determination kit using forceps.
! Ensure that the sample is completely submerged in the buoyancy liquid and
that no air bubbles have formed on the sample.
Max 220 g d
=
0.0001 g
4.4787
Weight in medium
END
+
Net
y The weighing display shows the step Weight in medium.
t To save the weight value, select v.
Max 220 g d
=
0.0001 g
11.86
END
+
y The balance calculates the density of the sample and displays this value.
t To exit the function, select x.
y The application switches back to its original state. If the GLP printout function
is activated, the GLP printout is automatically printed (see “Printout” on
page31).
There is an online animation for this application. Click here to view it: