Specifications

PR 5220 Operating Manual Fieldbus Interface
Sartorius 87
7.3.6 Register 5: Device Type and Software Release
Only reading is permitted.
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
TYPE MSB
Byte 1
TYPE LSB
Byte 2
MAINVERSION
Byte 3
SUBVERSION
E.g. 5220 Rel 1.23 = 52200123
hex
7.3.7 Register 6: Board Number
Only reading is permitted.
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Board number MSB
Byte 1
""
Byte 2
""
Byte 3
Board number LSB
E.g. 148388723 = 08D83B73
hex
7.3.8 Register 7: (Reserved)
7.3.9 Register 8 ...15: Weight Data
Only reading is permitted.
The gross, net and tare weight are stored as DINT fixpoint.
The real data value is derived from DINT and EXPO as follows :
Value
Real
=reading
DINT
*10
(-EXPO
Register 8
Current gross value
Register 9
Current net value, if tared; otherwise gross
Register10
Current tare value, if tared, otherwise 0
Register11
Value on the front-panel display
Register12
Reserved
Register13
Reserved
Register14
Max (FSD)
Register15
Reserved (free)
7.3.10 Register 22 ... 27: Limit Values (Read/Write)
Register 22 Limit 1 on
Register 23 Limit 1 off
Register 24 Limit 2 on
Register 25 Limit 2 off
Register 26 Limit 3 on
Register 27 Limit 3 off