Specifications

PR 5220 Operating Manual SMA Protocol
Sartorius 73
6.3.4 Scale Data
6.3.4.1 Scale Data – First Line
Comman
d:
<LF>A<CR>
Reply: The scale sends the first line of its scale data.
<LF><SMA>:<yyyyyy><CR>
For details, see Chapter 6.4.1 .
6.3.4.2 Scale Data – Other Lines
Command: <LF>B<CR>
Reply: The scale sends additional lines of its scale data.
<LF><MFG>:<yyyyyy><CR>
For details, see Chapter 6.4.1 .
6.3.5 Scale Information
6.3.5.1 Scale Information - First Line
Command: <LF>I<CR>
Reply: The scale sends the first line of its scale information.
<LF><SMA>:<yyyyyy><CR>
For details, see Chapter 6.4.1 .
6.3.5.2 Scale Information – Other Lines
Command: <LF>N<CR>
Reply: The scale sends additional lines of its scale information.
<LF><TYP>:<yyyyyy><CR>
For details, see Chapter 6.4.1 .
6.3.6 Escape Command
Command: <ESC>
Reply: This is the only command which does not work according to the <LF>c<CR> protocol
principle. It does not have a reply. The <ESC> character is detected at any time and cancels
any current command.
6.4 SMA Reply Messages
In this section, the replies are described in detail. The data format of each reply has a fixed length. The
communication error is the only exception from this pre-definable format. Thus the controlling computer
can check each reply according to fixed rules, because each data field is in a fixed position.