Specifications

SMA Protocol PR 5220 Operating Manual
72 Sartorius
6.3.2.2 Request for Taring with Fixtare Value
Command: <LF>T<xxxxxx.xxx><CR>
Reply: The scale makes a taring attempt using the fixtare value <xxxxxx.xxx> and signals the tared
status in the <s> and <n> characters.
<LF><s><r><N><m><f><xxxxxx.xxx><uuu><CR>
For details, see Chapter 6.4.1 .
6.3.2.3 Request for Zero Setting of the Scale
Command: <LF>Z<CR>
Reply: The scale makes a zero setting attempt and signals the zero status in the <s> character.
For this function, the stability condition must be met. The maximum waiting time for stability
is set under [Tare timeout]; see Chapter 5.4.13.
<LF><Z><r><n><m><f><xxxxxx.xxx><uuu><CR>
For details, see Chapter 6.4.1 .
Unless the scale is in the zero setting range, an error reply is generated.
6.3.2.4 Request for Tare Resetting
Command: <LF>C<CR>
Reply: The scale deletes the tare weight and signals the tare reset status in the <n> status character.
The scale tare is reset.
<LF><s><r><G><m><f><xxxxxx.xxx><uuu><CR>
For details, see Chapter 6.4.1 .
6.3.3 Scale Diagnosis
Command: <LF>D<CR>
Reply: The scale starts the diagnosis and returns a diagnosis reply.
<LF><r><e><c><m><CR>
For details, see Chapter 6.4.1 .