Specifications

PR 5220 Operating Manual Commissioning
Sartorius 59
5.8.1 Adapting the Analog Output
The output current can be adapted in small ranges. This is required, if small deviations from the nominal
value occur in a connected PLC.
Open the menu with
[Show HW-slots]-[Slot 1/2 PR5510/07 analog I/O]:
Info/HW-Slots
Builtin RS 232
Slot 1 PR 5510/04 analog I/O
PR 5510/07 is fitted in Slot 1
Slot 2 -empty-
Slot 3 Builtin Digital I/O
Slot 4 Profibus-DP
Select [PR 5510/07 analog I/O] and confirm with
Info/HW-Slots
PR5510/07 on slot 1
In use by PLC task 2
Analog output 0.0000 mA
Select with
Stop PLC Stop I/O Adjust Reset
Press [Adjust]
Adapt analog output ?
Safety prompt: Reply [Yes] to start adapting.
Info/HW-Slots/Adjust Analog Output Slot 1
Output 4.000 mA
Measured 4.004 mA
Enter e.g. the value for 4 mA measured by the connected PLC under [Measured].
After pressing
, the 2nd value (20 mA) is displayed:
Info/HW-Slots/Adjust Analog Output Slot 1
Output 20.000 mA
Measured 20.010 mA
Enter e.g. the value for 20 mA measured by the connected PLC under [Measured].
Save settings ?
Reply [Yes] to save the data,
or [No] to keep the previous values.
If you want to return to the factory settings (4 mA and 20 mA):
Press [Reset]: A safety prompt is displayed:
Reset to default ?
Reply [Yes] for reset to the factory settings, or
[No], if you want to keep the entered values unchanged.