Specifications
PR 5220 Operating Manual Commissioning
Sartorius 55
2. Determining a condition
The selected action of each digital input can be combined with a condition that must be met for signal
change from 0 to 1 (on) or for signal change from 1 to 0 (off). Select the condition from the list; see Chapter
5.7.1. No condition is defined when selecting [no condition]; the action is executed directly.
Example: Taring via digital input, only if the instrument is set to gross:
Setup/Digital I/O parameter
Output 1 marker bit 1 X64=1
Output 2 marker bit 2 X65=1
Output 3 marker bit 3 X66=1
Input 1 on set tare X113=1
Condition
tare active
X58=0
Input 1 off -no action-
Input 2 on -no action-
Input 2 off -no action-
Input 3 on -no action-
Input 3 off -no action-
When input 1 changes from 0 to 1 [Input 1 on], a taring signal is activated only, if the condition under
[Condition] is met (here: the scale was not tared previously; this is determined by [tare active X58=0]).