Specifications
Commissioning PR 5220 Operating Manual
52 Sartorius
5.7 Limit Values, Digital Inputs and Outputs
5.7.1 Conditions for Limit Values and Digital Inputs, States for Outputs
The limit values and digital inputs can be combined with conditions that must be met before an action is
executed. With outputs, these conditions can be output as states directly. Selection is from the following
list:
Selection list for [condition]
Function SPM bit(dec)
Explanation
Condition is met, if not true (=0) or if true (=1)
no condition ----- No condition
actual diginp1 X00=0/1 Digital input 1
actual diginp2 X01=0/1 Digital input 2
actual diginp3 X02=0/1 Digital input 3
actual limit 1 X16=0/1 Limit signal 1
actual limit 2 X17=0/1 Limit signal 2
actual limit 3 X18=0/1 Limit signal 3
ADC error X32=0/1 Error in load cell/ ADC area, negative/ far outside scale
above Max X33=0/1 Weight above Max
overload X34=0/1 Weight above (Max plus the range which is still permissible)
below zero X35=0/1 Weight below zero
center zero X36=0/1 Within 1/4 d of zero
inside ZSR X37=0/1 Within zero-setting range
standstill X38=0/1 Mechanical stability of the scale
dimmed X39=0/1 Weight below zero or above Max
command error X48=0/1 Command was no executed; e.g. zero-setting command, but the
weight is out of the zero-setting range
command busy X49=0/1 Command is being executed
power fail X50=0/1 Set after power-on (=power failure)
test active X56=0/1 Analog test was started
cal active X57=0/1 Calibration was started
tare active X58=0/1 Tare step is active
marker bit 1 X64=0/1 Marker bit 1; after power-on, the markers are set to '0'
marker bit 2 X65=0/1 Marker bit 2,
marker bit 3 X66=0/1 Marker bit 3,