Specifications
Commissioning PR 5220 Operating Manual
48 Sartorius
5.6.2 SMA Protocol
Select an RS-485 interface under
-[Serial ports parameter].
Setup/Serial ports
Remote display none
SMA
Builtin RS485
Param
Only the baud rate is adjustable; the other parameters are fixed.
Setup/Serial ports/Builtin RS485
Assigned to SMA
Baudrate
9600 bd
Press
to select [Baudrate] and
Bits 8
set the baud rate with
Parity none
Stopbits 1
The SMA protocol is described in Chapter 5.7.