Specifications
Commissioning PR 5220 Operating Manual
44 Sartorius
WP A/Calibration
Tare timeout 2.5 s
Zeroset range 50.00 d
Zerotrack range 0.25 d
Zerotrack step 0.25 d
Zerotrack time 0.0 s
Overload 9 d
[Tare timeout]
Enter a timeout value between 0.1 and 25 s for a taring/zero set command that cannot be executed (e.g., if
scale mechanically unstable, filter settings faulty, resolution too high, stability condition too narrow).
[Zeroset range]
Determine a +/- range around the zero point determined by the dead load during calibration; within this
range:
- the displayed gross weight can be set to zero by pressing the zero-setting key (or by a corresponding
external command), and
- automatic zero tracking is active
Available range: 0.00 d to 10000.00d
In 'legal-for-trade' mode a value <= 2 % of Max. must be entered. Example: 60 d for 3000e, class III.
[Zerotrack range]
Range within which automatic zero tracking compensates deviations; 0.25 to 10000.00d. In 'legal-for-trade'
mode a value of 0.25 d has to be entered.
[Zerotrack step]
Step for automatic tracking; 0.25 to 10d. In 'legal-for-trade' mode a value of 0.25 d has to be entered.
[Zerotrack time]
Enter a time interval for automatic zero tracking within 0.0 (tracking switched off) and 25 s. In 'legal-for-
trade' mode a value of 0.25 d has to be entered.
[Overload]
Weighing range above the maximum capacity (Max), without error message. Available range: 0 to
9999900 d. In 'legal-for-trade' mode max. 9 d=e has to be entered.
WP A/Calibration
Min
50 d
Range mode
Single range
[Min]
Minimum weight at which a print command can still be executed. Range is 0 to 9999900 d.
In 'legal-for-trade' mode min. 20 d has to be entered.