Specifications

PR 5220 Operating Manual Commissioning
Sartorius 43
5.4.13 Parameter Input
Open the menu via
-[Calibration]-[Param].
WP A/Calibration
Measuretime
160 ms
Digital filter bessel
External supply 8 - 12 VDC
Fcut 3.00 Hz
This line is shown only if the filter is on.
Test mode absolute
W & M
none
Standstill time 0.50 s
Standstill range 1.00 d
[Measuretime]
Select the measuring time; possible values: 5, 10, 20, 40, 80, 160, 320, 640, 960, 1280, 1600 ms. In 'legal-
for-trade' mode select <= 1 s.
[Digital filter]
The digital filter can be switched on only with the measuring time set to <= 160 ms.
Select the filter characteristic [off], [bessel], [aperiod.], [butterw.], [tscheby.].
[External supply]
With external load cell supply (e.g. 7.5 VDC via 1626/60 with MX8 = closed), [below 8 VDC] can be selected
to adapt the sense voltage monitoring to the lower supply voltage.
[Fcut]
Enter the cut-off frequency for the filter (0.1 – 80 Hz); the setting is dependent on the measuring time.
[Test mode]
With [Absolute], the test value is calculated when the test is called. With [relative], the deviation from the
initially stored test value is displayed; see Chapter 5.4.11.
[W & M]
Setting for 'legal-for-trade' mode; select [none], [OIML], [NTEP] (for USA) or [NSC] (for Australia), see
Chapter 5.4.13.1.
[Standstill time]
Settings [Standstill range] and [Standstill time] are required for determining the mechanical stability of the
scale. Input in seconds; permissible range 0.01 bis 2 sec. (max. measuring time x 32). If 0 is set, stability is
not checked. The stability time must not be less than the measuring time and not greater than 32 times the
measuring time.
[Standstill range]
The scale is stable as long as any changes in the weight value are within this range; permissible range: 0.01d
to 10.00d. In 'legal-for-trade' mode select <= 1 d.