Specifications

Commissioning PR 5220 Operating Manual
38 Sartorius
5.4.6 Calibration with Weight (by Load)
Select [by load] for calibration using weight.
Place CAL weight
on the scale
and enter value
The weight value for the calibration weight must be entered in a separate
window.
2000 kg
After applying the weight, enter the weight value and confirm with
. The weight unit for the calibration
weight (press
to change) may differ from the unit in the instrument; conversion is automatic.
Afterward, the following message is displayed:
Setting SPAN by load
Weight value, weight unit and measuring signal in mV/V corresponding to this value are displayed in the
[Calibrated at] line.
WP A/Calibration
Max 3000 d 3000 kg
Scale interval 3000 d 1 kg
Deadload at 165.11 kg 0.057920
Max at 3000.00 kg 1.052369
Calibrated at 2000 kg 0.701579
Sensitivity 876.97
4.209600
by load by mV/V by data Linear. Test
Set SPAN failed
No stability
The scale is not stable.
Remedial action: Check the mechanical function of the scale; adapt the
filter setting; reduce the resolution; if necessary, adapt the stability
conditions.
Set SPAN failed
Load below deadload
The weight on the scale is less than the dead load after input of the
weight value.
The next step is calculation of the test value with [Test] (see Chapter 5.4.11), and calibration is completed
with
(see Chapter 5.4.12).
Determining span without weight
WP A/Calibration
Max 3000 d 3000 kg
Scale interval 3000 d 1 kg
Deadload at 3.00 kg 0.001000
Max at 3000.00 kg 0.000000
Not calibrated
Sensitivity 833.33
4.000000
by load by mV/V by data Linear. Test