Specifications

Commissioning PR 5220 Operating Manual
34 Sartorius
5.4.2 Selecting the Calibration Mode
You can choose between [New] and [Modify] with the softkeys:
New Modify Param
5.4.2.1 New Calibration
Open the menu via -[Calibration].
When you press [New], the data is set to default first and calibration is started.
SPAN and deadload
will be reset
You are prompted to confirm.
Press [Continue] for the default settings, or [Cancel] to cancel the selection.
Default settings with [New]:
WP A/Calibration
Max 3000 d 3000 kg
Scale interval 3000 d 1 kg
Deadload at 0.000000
Max at 1.000000
Not calibrated
Sensitivity
833.33
4.000000
Test
5.4.2.2 Changing the Calibration
[Modify] may be used only for minor changes (e.g. changing the dead load, adapting mV/V values
for dead load and/or Max); otherwise, always use [New].
Open the menu via
-[Calibration]-[Modify].
WP A/Calibration
Max 3000 d 3000 kg
Scale interval 3000 d 1 kg
Deadload at 1.07 kg 0.000358
Max at 3000.00 kg 1.000000
Calibrated at 3000.00 kg 1.000000
Sensitivity
833.33
4.000000
by load by mV/V Test
For setting a new value for Dead load, press
to select [Deadload] and either enter a new value
with [by mV/V] or discharge the scale/hopper and press [by load].
Exit calibration
without CalcTest?
When closing the menu with
you are prompted whether the menu
should be closed without calculation of the test value:
Reply [Yes] to close the menu.