Specifications
Installing the Instrument PR 5220 Operating Manual
24 Sartorius
3.2.7 DeviceNet Interface (PR 5220/04 only)
It is a complete DeviceNet adapter (SLAVE) with CAN controller and transfer rates of up to 500 kbit/s.
Connection:
5-contact screw terminal block (plug-in)
Transfer rate:
125, 250 and 500 kbit/s
Topology:
Parallel bus
Protocol:
DeviceNet master/slave
Polling method (polled I/O)
CRC error detection
to IEC 62026 (EN50325)
Max. 64 station nodes
Max. data width 512 bytes input &
output
Transport:
EIA RS-485
Configuration:
EDS file
MAC-ID (1…62)
Certificates/
conformity:
Compatible with DeviceNet specification
Vol 1: 2.0, Vol 2: 2.0
ODVA certificate in accordance with
conformity test software version A-12
Suitable for industrial applications to CE,
UL and cUL
Cable:
DeviceNet, color: petrol-green
2x2 shielded twisted pair
Cable impedance:
150 ohms
Bus termination:
120 ohms at the cable ends
Electrical isolation:
Yes, optocoupler and DC/DC converter
Connecting diagram for a master with three slaves