Safety Instructions
!
The seals affixed to this equipment indicate that only authorized service
technicians are allowed to open the equipment and perform maintenance work
so that safe and trouble-free operation of the equipment is ensured and the
warranty remains in effect.
Before opening the weighing cell or A/D converter enclosure, make sure to
unplug the weighing platform from the power supply!
Instructions for Use in Hazardous Areas
Weighing platforms of the series ISI 150IGG-H… and IS300IGG-H comply
with the European Standards EN 50 021: 1999 and EN 5028-1-1-1
according to the Statement of Conformity, No. TüV 01 ATEX 1739 X,
including the 1
st
and 2
nd
Supplements (see “TÜV Certificates”). Therefore, they
are suitable for use in Zones 2 and 22 hazardous areas. Their approval
marking is:
h II 3 GD IP67 Eex nR II T6 T80°C
The type of explosion protection used in these weighing platforms involves
protection of the housing against penetration of potentially explosive vapor. For
this purpose, IP67 protection of the platform must be ensured when it is
connected to the power source. Opening the platform or disconnecting cables
while the scale is plugged into the power supply is strictly forbidden. Please
read the installation instructions in the Annex 35751-000-16-A4. To install the
power supply, please follow the installation instructions given in the section
entitled “Connecting the Weighing Platform” in “Getting Started.”
If you will be operating the platform outside the European Community in a
Zone 2 hazardous area, you must comply with the national electrical code and
applicable safety regulations of your country. Please ask your local Sartorius
service technician, office or dealer for information on the currently valid
regulations applicable in your country.
The weighing platform may not be operated in Zone 0, 1, 20 or 21 hazardous
areas, as it does not have an EX approval certificate for these areas.
Any tampering with the weighing platform by anyone, other than installation of
preload devices, will result in forfeiture of all claims under the manufacturer's