Operating instructions

26
Problem Possible Causes Solution
The error code "ERR 12" is
displayed
You tried to store a value in the
tare memory for applications,
although this memory has been
blocked by menu code 2 2 x
You entered a numeric tare that
is not allowed
Check the configuration
– Check the value entered
The error code "ERR 17" is
displayed
Internal calibration is not
possible because the preload
setting is too high
See "Setting the Preload" in the section on the
operating menu
The error code "ERR 19" is
displayed
The remaining weighing
capacity available is too low
because the relatively high
preload setting has reduced the
maximum capacity
See "Setting the Preload" in the section on the
operating menu
The special code "A" remains
displayed
None of the keys has been
pressed since the weighing
platform was switched on
Press any key
The weight readout changes
constantly
Unstable environment (ambient
conditions; e.g., too much
vibration or exposure to
excessive drafts)
A foreign object is caught
between the load plate and the
platform frame
Set up the weighing platform in a different
area
Access the "Devices" setup menu to adapt the
weighing platform to the particular
environment
Remove the foreign object
The readout of the weighing
result is obviously wrong
The weighing platform has not
been calibrated/ adjusted
The weighing platform was not
tared before weighing
The weighing platform is tilted
Calibrate/ adjust the weighing platform
Tare before weighing
Level the weighing platform as described
in the instructions
If any other errors occur, please contact your nearest Sartorius service center or your dealer.