Operating instructions

21
IS 300 IGG-H0CE
Weighing range structure SuperRange SingleRange Multi-Interval
Weighing range Standard Configurable
Maximum capacity, Max kg
300 61 31 150
Readability g
2 1 1 0 ... 30 kg 10
30 ... 60 kg 20
60 ...150 kg 50
Zero tracking ± 2% kg
± 6 ± 1.22 ± 0.62 ± 3
Initial zero-setting range kg
±10 ±10 ±10 ±10
Preload (electronic)
1
) kg
60 120 150 30
ISO test function
2
) yes yes yes yes
Metrological specifications:
Accuracy class g
l l
Verification scale interval g 10 10 10/20/50
Minimum capacity, Min, g
acc. to CD³)
50 50 200
Temperature range °C 0 ... +40
Select particular group of
specifications
(use code)::
Spec. grp.1*
(10 3 1)
Spec. grp.2
(10 3 2)
Spec. grp.3
(10 3 3)
Spec. grp.4
(10 3 4)
Weighing range structure SingleRange
Weighing range
Configurable
Maximum capacity, Max kg
300 300 151 61
Readability g 50 100 50 20
Zero tracking ± 2% kg
± 6 ± 6 ± 3.02 ± 1.22
Initial zero-setting range kg
±10 ±10 ±10 ±10
Preload (electronic)
1
) kg
60 60 209 299
ISO test function ²) yes yes yes yes
Metrological specifications:
Accuracy class
l l l l
Verification scale interval g 50 100 50 20
Minimum capacity, Min, g
acc. to CD³)
1,000 2,000 1,000 400
Temperature range °C
0 ... +40
Select particular group of
specifications
(use code)::
Spec.grp.6
(10 3 6)
Spec. grp.7
(10 3 7)
Spec. grp.8
(10 3 8)
Spec. grp.9
(10 3 9)
1
) With relatively high preloads, the maximum capacity is reduced by 1-kg intervals
2
) Up to a max. preload of 120 kg
3) Council Directive 2009/23/EC on non-automatic weighing instruments used in the EU and the Signatories of the Agreement on the European
Economic Area
The weighing platform has a built-in calibration weight.
Should external calibration be necessary, you will need one of the following calibration weights:
– Readability of 1 g: 50 kg, accuracy class F1
– Readability of10 g or more: 50 kg, accuracy class M1
* factory setting