Specifications Charts
Access the configuration menu of the conected Sartorius terminal to set the code for the group of specifications of your
choice. Please note that when you change the configuration, this may also change the preload range besides the weigh-
ing capacity and the readability. Select the appropriate ID label and affix it to the weighing platform. For direc-tions,
refer to the section entitled "Selecting and Affixing the Respective ID Label." A verifiable scale consists of a weighing
platform (weighing module) and a display and control unit. The scale has the type designation with isoTEST HC BF or
SARTOCOWAT with HCBF : IS 150 IGG-H0CE
Weighing range structure SuperRange SingleRange Multi-Interval
Weighing range Standard Configurable
Maximum capacity, Max kg
150 61 31 150
Readability g
1 1 1 0 ... 30 kg 10
30 ... 60 kg 20
60 ...150 kg 50
Zero tracking ± 2% kg
± 3 ± 1.22 ± 0.62 ± 3
Initial zero-setting range kg
±10 ±10 ±10 ±10
Preload (electronic)
1
) kg
30 120 150 30
ISO test function
2
) yes yes yes yes
Metrological specifications:
Accuracy class g
KK
–
l
Verification scale interval g 10 10 – 10/20/50
Minimum capacity, Min, g
acc. to CD³)
50 50 – 200
Temperature range °C 0 ... +40
Select particular group of
specifications
(use code)::
Spec. grp.1*
(10 3 1)
Spec. grp.2
(10 3 2)
Spec. grp.3
(10 3 3)
Spec. grp.4
(10 3 4)
Weighing range structure SingleRange
Weighing range
Configurable
Maximum capacity, Max kg
150 150 61 31
Readability g 20 50 20 10
Zero tracking ± 2% kg
± 3 ± 3 ±1.22 ± 0.62
Initial zero-setting range kg
±10 ±10 ±10 ±10
Preload (electronic)
1
) kg
30 30 120 150
ISO test function ²) yes yes yes yes
Metrological specifications:
Accuracy class
l l l l
Verification scale interval g 20 50 20 10
Minimum capacity, Min, g
acc. to CD³)
400 1,000 400 200
Temperature range °C
0 ... +40
Select particular group of
specifications
(use code):
Spec.grp.5
(10 3 6)
Spec. grp.6
(10 3 7)
Spec. grp.7
(10 3 8)
Spec. grp.8
(10 3 9)
1
) With relatively high preloads, the maximum capacity is reduced by 1-kg intervals
2
) Up to a max. preload of 120 kg
3) Council Directive 2009/23/EC on non-automatic weighing instruments used in the EU and the Signatories of the Agreement on the European
Economic Area
The weighing platform has a built-in calibration weight. Should external calibration be necessary, you will need one of the following weights:
– Readability of 1 g: 50 kg, accuracy class F1
– Readability of 10 g or more: 50 kg, accuracy class M1
* factory setting