Operating instructions
Calibration/Adjustment
Preparation
● Select the scale function for “calibration/adjustment”: press H
● To select the
Balance/scale functions: press the O soft key
● Select
Calibration/adjustment: press the O soft key
Calibration/ CAL key function Ext. cal./adj.: default wt.
adjustment Ext. cal./adj.: user-defined wt.
ο Internal cal./adjustment
Key blocked
reproTEST*
Selection mode
Cal/adjustment Calibrate, then auto adjust
sequence ο Calibrate, then manual adjust
isoCAL function ο Off
Only adjustment prompt
On and reset application
On without resetting app.
Start automatic User-def. Adj. time1
adjustment adjustment times Adj. time2
Adj. time3
ο isoCAL
Print GMP adjustment ο Automatic if “GLP” is selected
record On request, from record memory
Parameter for Weight ID (Wt. ID):
external weight Cal./adj. wt.:
ο = factory setting
● Save settings and exit Setup menu: press the
oo soft key
44