Operating instructions
4
Tables of Specifications
Model IS06BBE-SX IS2CCE-SX IS6CCE-HX IS6CCE-SX IS12CCE-
SX
Readability g 0.001 0.01 0.01 0.1 0.1
.........
Weighing capacity g 620 2200 6200 6200 12,000
Max. overload capacity kg 3 12 12 25 25
Tare range (subtractive) g – 620 – 2200 – 6200 – 6200 – 12000
Preload (can be electronically g 93 110 – 1240 1200
compensated without reducing the weighing capacity)
Max. preload at start of isoTEST g 110 1300 5200 6400 10000
calibration/adjustment
(platform must be zeroed)
Repeatability <±g 0.001 0.01 0.01 0.05 0.05
Linearity <±g 0.002 0.02 0.02 0.1 0.2
Sensitivity drift within +10 to +30 °C <±/K 2 • 10
-6
2 • 10
-6
2 • 10
-6
4 • 10
-6
4 • 10
-6
Response time (average) s 1.5 1.5 1.5 1 1
External calibration weight g 500 (E2) 2000 (F1) 5000 (E2) 5000 (F2) 5000 (F1)
(of at least accuracy class...)
Load plate dimensions mm Ø 130 218 x 200 218 x 200 218 x 200 218 x 200
Platform dimensions (W x D x H) mm 240x294x86 240x294x86 240x294x86 240x294x86 240x294x86
Net weight, approx. kg 7 7.3 8.4 7.3 7.3
Dust and water protection rating for the housing acc. to EN 60529 IP54
AC power source Power supply model YPS02-X or AC adapter model YPS02-Z (100-240V)
(must be connected by the weighing equipment dealer or a Sartorius service technician)
Frequency Hz 48 – 62
Allowable ambient operating temperature 0 °C to +40 °C (273 °K to 313 °K; 32 °F to 104 °F)
Operating temperature range +10 °C to +30 °C (or 0 °C to +40 °C; depends on configuration)
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display update (depends on filter level selected) 0.1 - 0.1 - 0.2 - 0.4 (normal output rate)
Power consumption Average: 25VA
Selectable weight units: Grams, kilograms, carats, pounds, ounces, Troy ounces, Hong Kong taels,
Singapore taels, Taiwanese taels, grains, pennyweights, milligrams, parts per
pound, Chinese taels, mommes, Austrian carats, tola, baht and mesghal
Built-in RS-485 interface (standard equipment; xBPI protocol):8 bits, odd parity; transmission rate: 9600 -
38,400 baud; half-duplex; convertible to RS-232 (SBI protocol): 7 bits, Parity: even,
odd, mark, space; transmission rates: 150 — 19,200 baud; 1 or 2 stop bits, software/
hardware handshake; factory settings: 1200 baud, odd parity, 1 stop bit, hardware
handshake with 2 characters after CTS
Model series Marking (ATEX) Use in IP rating Markings in acc. with FMRC Use in
Zone approval for USA and CSA
approval for Canada
IS.. BBE-.X.. 1, 2 (gas) IP 54 IS CL I, DIV 1, GR A,B,C,D, T4 Class I, Division 1 (gas)
h II 2 G EEx ib IIC T4 CL l, ZONE 1 AEx ib IIC T4 Class I, Zone 1 (gas)
IS.. CCE-.X.. 1, 2 (gas) IP 54 IS CL I, DIV 1, GR A,B,C,D, T4 Class I, Division 1 (gas)
h II 2 G EEx ib IIC T4 CL l, ZONE 1 AEx ib IIC T4 Class I, Zone 1 (gas)
IS… EDE-.X.. 1, 2 (gas) IP 65 IS CL I, DIV 1, GR A,B,C,D, T4 Class I, Division 1 (gas)
h II 2 G EEx ib IIC T4 CL l, ZONE 1 AEx ib IIC T4 Class I, Zone 1 (gas)
20, 21, 22 IP 65 IS CL II,III, DIV 1, GR E,F,G, T4 Class II,III Division 1
h II 1 D T135°C (dust) (dust+ fibers)
IS… IGG-HX 1, 2 (gas) IP 65 and IP67 IS CL I, DIV 1, GR A,B,C,D, T4 Class I, Division 1 (gas)
h II 2 G EEx ib IIC T4 CL l, Zone 1 AEx ib IIC T4 Class I, Zone 1 (gas)
h II 1 D T135°C 20, 21, 22 IP 65 and IP 67 IS CL II,III, DIV 1, GR E,F,G, T4 Class II,III Division 1
(dust) (dust + fibers)
* = factory setting