User manual
8 Cubis MSU Brief Instructions
Information for the User
Touch Screen User Interface
Many of the operating steps involve use of the display and softkeys. The layout
of the display depends on whether the application software is in operating mode
or whether a menu is open (configuration mode).
Display in Operating Mode
1 Function area: Shows the active application (here: step 1 of 2)
2 Line for metrology data with weighing capacity (left) and reading precision
(resolution) of the balance.
3 Value area with the current measured value
4 Area for status messages and alerts (see below)
5 Task area with instructions on how to proceed
6 Function bar with the currently available softkey functions.
7 Softkeys: Associated with the function indicated above them or used for
navigation
8 User info area: Displays the current user, date and time
9 Scaled display of measured value (percentage of weighing capacity utilization)
Status Messages and Alerts
The following messages may be shown here:
isoCAL shown continuously: isoCAL function active (extended temperature range)
isoCAL flashing: Calibration/adjustment required
Level me flashing: Balance not level, leveling required
Print: Printout is active; print job not yet completed
GLP: GLP-compliant printing is active (includes header)
SQmin: SQmin value is shown if the weight value is less than SQmin
U or U* or PA: The selected DKD value (measurement uncertainty) is shown